Mechanical reliability of the Jarvik 2000 Heart.

Michael P Siegenthaler1, O H Frazier, Friedhelm Beyersdorf, Jürgen Martin, Hillel Laks, John Elefteriades, Asghar Khaghani, Ulf Kjellman, Bansi Koul, John Pepper, Robert Jarvik, Stephen Westaby.   

Abstract

BACKGROUND: Device failure is a limitation of permanent mechanical circulatory support. We studied the mechanical reliability of the Jarvik 2000 Heart, an axial flow pump with ceramic bearings designed to provide more than 10 years' durability.
METHODS: The Jarvik 2000 Heart was implanted in 102 patients between April 2000 and December 2004. Eighty-three pumps with an abdominal driveline were implanted as a bridge-to-transplantation, and 19 with postauricular power supply as lifetime therapy. Eighteen pumps were recovered intact after clinical use and run continuously on the bench to further assess durability.
RESULTS: No implantable component failure occurred either in patients or during bench testing. The cumulative pump run-time was 110 years: 59 years overall in vivo and 51 years in vitro. The mean support time for bridge-to-transplant recipients was 159 days, and for discharged lifetime-therapy recipients 551 days. Six recipients were supported moer than 2 years, with the longest ongoing approaching 5 years. External cables caused three system failures, with a 95% freedom from system failure at 4 years. Device malfunctions, related to external cables (9) and lack of a backup battery (1), caused no adverse consequences. Before introduction of noncorrosive, gold-plated stainless steel connectors, corrosion was observed on three connectors to the retroauricular power supply.
CONCLUSIONS: The Jarvik 2000 Heart has had no implantable component failure. Meaningful durability data and failure mode can only be established by real-time testing in patients. The reliability and dependability of this device, in addition to the exchangeability of external components, give promise for long-term circulatory support in critically ill heart failure patients.

Entities:  

Mesh:

Year:  2006        PMID: 16631667     DOI: 10.1016/j.athoracsur.2005.12.013

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  9 in total

1.  Therapeutic hypothermia after global cerebral ischemia due to left ventricular assist device malfunction.

Authors:  Shunsuke Saito; Koichi Toda; Shigeru Miyagawa; Yasushi Yoshikawa; Hiroki Hata; Daisuke Yoshioka; Satoshi Kainuma; Shohei Yoshida; Yoshiki Sawa
Journal:  J Artif Organs       Date:  2019-03-02       Impact factor: 1.731

Review 2.  A ventricular assist device as a bridge to recovery, decision making, or transplantation in patients with advanced cardiac failure.

Authors:  Siyamek Neragi-Miandoab
Journal:  Surg Today       Date:  2012-07-20       Impact factor: 2.549

3.  Long-term continuous-flow left ventricular assist devices (LVAD) as bridge to heart transplantation.

Authors:  Matteo Pozzi; Raphaël Giraud; Piergiorgio Tozzi; Karim Bendjelid; Jacques Robin; Philippe Meyer; Jean François Obadia; Carlo Banfi
Journal:  J Thorac Dis       Date:  2015-03       Impact factor: 2.895

4.  Clinical results with Jarvik 2000 axial flow left ventricular assist device: Osaka University Experience.

Authors:  Daisuke Yoshioka; Goro Matsumiya; Koichi Toda; Taichi Sakaguchi; Yasushi Yoshikawa; Shunsuke Saito; Hikaru Matsuda; Yoshiki Sawa
Journal:  J Artif Organs       Date:  2014-07-22       Impact factor: 1.731

Review 5.  Left ventricular assist devices as a bridge to cardiac transplantation.

Authors:  Christopher T Holley; Laura Harvey; Ranjit John
Journal:  J Thorac Dis       Date:  2014-08       Impact factor: 2.895

6.  Jarvik 2000 with postauricular cable as destination therapy: first clinical case in Japan.

Authors:  Keitaro Domae; Koichi Toda; Ryohei Matsuura; Shigeru Miyagawa; Yasushi Yoshikawa; Hiroki Hata; Shunsuke Saito; Daisuke Yoshioka; Fusako Sera; Kei Nakamoto; Satoru Oshino; Haruhiko Kishima; Yasushi Sakata; Yoshiki Sawa
Journal:  J Artif Organs       Date:  2019-09-13       Impact factor: 1.731

7.  Inappropriate use of neurally adjusted ventilator assist.

Authors:  Francesca Iodice; Giuseppina Testa; Antonio Amodeo; Paola Cogo
Journal:  BMJ Case Rep       Date:  2012-09-03

8.  Current trends in implantable left ventricular assist devices.

Authors:  Jens Garbade; Hartmuth B Bittner; Markus J Barten; Friedrich-Wilhelm Mohr
Journal:  Cardiol Res Pract       Date:  2011-05-09       Impact factor: 1.866

Review 9.  Mechanical Ventricular Assistance as Destination Therapy for End-Stage Heart Failure: Has it Become a First Line Therapy?

Authors:  Massimo Bonacchi; Guy Harmelin; Marco Bugetti; Guido Sani
Journal:  Front Surg       Date:  2015-08-03
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.