Irregularities on the surface of single-piece AcrySof SA60AT intraocular lenses.

Dan Q Nguyen1, Tarek A Saleh, Suresh K Pandey, Adrian K Bates.   

Abstract

PURPOSE: To report a case series of what initially appeared to be macroscopic biomaterial deposits and irregularities on the surface of AcrySof SA60AT intraocular lenses (IOLs) (Alcon Inc.)
SETTING: Department of Ophthalmology and Somerset Hospital, Taunton, United Kingdom; Department of Ophthalmology and Visual Sciences, John A Moran Eye Center, University of Utah, Salt Lake City, Utah, USA; and Intraocular Implant Unit, Sydney Eye Hospital, University of Sydney, Sydney, Australia.
METHODS: Twelve AcrySof SA60AT IOLs were initially thought to have irregularities on their surfaces when examined under the operating microscope before implantation in the capsular bag during cataract surgery. The IOLs were sent for further analysis, including gross examination and light and scanning electron microscopy, to John A. Moran Eye Center, Salt Lake City, Utah, USA, and to Alcon Laboratories, Hemel Hempstead, United Kingdom.
RESULTS: Light microscopy revealed the presence of irregular fine lesions on the IOLs' optics and haptics. The lesions consisted of elevations and depressions of the IOL surface that were present on both the anterior and the posterior surfaces of the IOLs. Scanning electron microscopy confirmed the presence of physical/mechanical damage to the IOLs, which was thought to originate from the packaging. The manufacturer believed this to have resulted from compression of the IOL between the container and its cover.
CONCLUSIONS: Implantation of an imperfect IOL remains a potentially serious occurrence. Faults still occur in modern IOLs, and some defects can be detected by examining the IOL under the operating microscope before implantation into the eye.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16631064     DOI: 10.1016/j.jcrs.2005.12.074

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  4 in total

1.  Optic surface changes in Intraocular lens scaffold: An ex vivo study.

Authors:  Dhivya Ashok Kumar; Amar Agarwal; Dhanalakshmi Panneerselvam; Josheela Sheebakalyani Gnanaprakasam
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-05-28       Impact factor: 3.117

2.  Preloaded and non-preloaded intraocular lens delivery system and characteristics: human and porcine eyes trial.

Authors:  Byunghoon Chung; Hun Lee; Moonjung Choi; Kyoung Yul Seo; Eung Kweon Kim; Tae-Im Kim
Journal:  Int J Ophthalmol       Date:  2018-01-18       Impact factor: 1.779

3.  Evaluation of semi-preloaded intraocular lens delivery system.

Authors:  Rajesh Rajesh
Journal:  Rom J Ophthalmol       Date:  2019 Jan-Mar

4.  Clinical Evaluation of a Novel Preloaded Intraocular Lens Delivery System During Routine Cataract Surgery.

Authors:  Daniel Black; Dean Corbett; Timothy V Roberts; Brendan Cronin; Pamela J Smith; D Priya Janakiraman; Beth E Jackson
Journal:  Clin Ophthalmol       Date:  2020-08-12
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.