Passive electrode effect reduces defibrillation threshold in bi-filament middle cardiac vein defibrillation.

John R Paisey1, Arthur M Yue, Frederick Bessoule, Paul R Roberts, John M Morgan.   

Abstract

AIMS: To investigate whether a passive electrode effect decreases defibrillation threshold (DFT) in multi-filament middle cardiac vein (MCV) defibrillation. METHODS AND
RESULTS: Twelve pigs underwent active housing (AH) insertion, with defibrillation coils placed transvenously in right ventricular apex and superior vena cava. The MCV was cannulated, and 1.12F, 50 mm coil electrodes (Ela Medical SA, France) were deployed in its right and left branches. Lead placement was possible in 11 of 12 animals. DFT (J, mean +/- SD) was determined by three-reversal binary search and compared between the MCV monofilament (single filament deployed) and the AH (25.9 +/- 10.9) and the MCV mono + passive filaments (both filaments deployed, one connected) and the AH (19.9 +/- 11.4); 24% DFT reduction P = 0.008.
CONCLUSION: A bystander electrode adjacent to a monofilament electrode in the MCV reduces DFT by 24% when compared with monofilament MCV alone. Microfilament electrodes decrease DFT as auxiliary anode but not as sole anode.

Entities:  

Mesh:

Year:  2006        PMID: 16627420     DOI: 10.1093/europace/euj034

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  1 in total

1.  Incidence and clinical predictors of low defibrillation safety margin at time of implantable defibrillator implantation.

Authors:  Zhongwei Cheng; Mintu Turakhia; Ronald Lo; Anurag Gupta; Paul C Zei; Henry H Hsia; Amin Al-Ahmad; Paul J Wang
Journal:  J Interv Card Electrophysiol       Date:  2012-03-06       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.