| Literature DB >> 16624545 |
A V Nabok1, A Tsargorodskaya, A Holloway, N F Starodub, O Gojster.
Abstract
A sensitive optical method of total internal reflection ellipsometry (TIRE) in conjunction with immune assay approach was exploited for the registration of T-2 mycotoxin in a wide range of concentrations from 100 microg/ml down to 0.15 ng/ml. Association constants of 1.4x10(6) and 1.9x10(7)mol(-1)s for poly- and monoclonal T-2 antibodies, respectively, were evaluated from TIRE kinetic measurements. According to TIRE data fitting, binding of T-2 molecules to antibodies (at saturation) has resulted in the increase in adsorbed layer thickness of 4-5 nm. The QCM impedance measurements data showed anomalously large mass increase and film softening, most likely, due to the binding of large T-2 aggregates to antibodies.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16624545 DOI: 10.1016/j.bios.2006.03.010
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618