Abstract

Arrestee urine specimens (930) were tested with DRI, CEDIA, and REMEDi; those that tested positive for amphetamines and opiates (616 and 414, respectively) were then confirmed by gas chromatography-mass spectrometry. The performance characteristics of these three preliminary systems were evaluated using the following commonly used parameters: true positive, true negative, false positive, and false negative. The sensitivity, specificity, and efficiency of these methods were also calculated. Data derived from this study indicated DRI and CEDIA adapted by this study generated acceptable preliminary test results for amphetamine/methamphetamine and morphine/codeine, but not for MDA/MDMA and REMEDi has lower sensitivity than DRI and CEDIA, but with better specificity and efficiency, supporting its use under emergency room settings where drug concentrations in overdose cases are expectedly at high levels.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16620534     DOI: 10.1093/jat/30.1.61

Source DB:  PubMed          Journal:  J Anal Toxicol        ISSN: 0146-4760            Impact factor:   3.367


  2 in total

Review 1.  Endogenous opiates and behavior: 2006.

Authors:  Richard J Bodnar
Journal:  Peptides       Date:  2007-09-11       Impact factor: 3.750

2.  Methotrimeprazine-induced corneal deposits and cataract revealed by urine drug profiling test.

Authors:  Seong Taeck Kim; Jae Woong Koh; Joon Mo Kim; Won Young Kim; Gwang Ju Choi
Journal:  J Korean Med Sci       Date:  2010-10-26       Impact factor: 2.153

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.