| Literature DB >> 16616252 |
Olivier H J Koning1, Jan-Willem Hinnen, Jary M van Baalen.
Abstract
Complete removal of an aortic endograft with suprarenal fixation is difficult. We report the use of a simple device to extract a Zenith endograft. This device is made by cutting off the tip of the cylinder of a 20-mL syringe and rounding off the edges. The main body of the graft is resheathed by advancing the cylinder cranially while keeping the graft in a stable position. This way, the graft collapses and the hooks are withdrawn without tearing the wall of the aorta. The supraceliac clamping time in our patient was less than 2 minutes. This technique offers a safe and fast solution to the potentially hazardous removal of an aortic endograft with suprarenal fixation.Entities:
Mesh:
Year: 2006 PMID: 16616252 DOI: 10.1016/j.jvs.2005.12.011
Source DB: PubMed Journal: J Vasc Surg ISSN: 0741-5214 Impact factor: 4.268