A clinical investigation of the fit of removable partial dental prosthesis clasp assemblies.

Daniel Dunham1, James S Brudvik, W Jack Morris, Kevin D Plummer, Stephen M Cameron.   

Abstract

STATEMENT OF PROBLEM: Removable partial denture frameworks are complex castings fabricated from high-shrinkage alloys. This may result in difficulty in achieving accurate fit.
PURPOSE: The purpose of this study was to evaluate which component(s) of the clasp assembly contact the abutment tooth clinically.
MATERIAL AND METHODS: New and existing frameworks with at least 1 conventional prepared occlusal rest were evaluated for fit. The spacing between the bottom of each of the rests of the 50 clasp assemblies and the corresponding rest seat were recorded with vinyl polysiloxane. The fit of each rest was determined with the use of a dial caliper by measuring, in micrometers, the thickness of the record between the rest and the bottom of the rest seat. The Kennedy classification was also recorded. A 2-sample t test was used to evaluate the difference in fit between tooth-tissue supported and tooth-supported designs (alpha=.05).
RESULTS: The average space between the rest and prepared rest seat was 193 +/- 203 microm, with a range of 0 to 828 mum. Twenty tooth-tissue frameworks had an average space of 136 +/- 160 microm and 30 tooth-supported frameworks had an average space of 230 +/- 222 microm. The 2-sample t test showed no significant difference for fit between tooth-tissue supported and tooth-supported frameworks (P=.1081). Twenty-four percent of rests had contact in the prepared rest seat.
CONCLUSION: The majority of rests evaluated did not contact the intended surfaces.

Entities:  

Mesh:

Year:  2006        PMID: 16616131     DOI: 10.1016/j.prosdent.2006.02.001

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  3 in total

1.  [Evaluation of one-piece polyetheretherketone removable partial denture fabricated by computer-aided design and computer-aided manufacturing].

Authors:  X X Li; Y S Liu; Y C Sun; H Chen; H Q Ye; Y S Zhou
Journal:  Beijing Da Xue Xue Bao Yi Xue Ban       Date:  2019-04-18

Review 2.  A Scoping Review on the Accuracy of Fit of Removable Partial Dentures in a Developing Digital Context.

Authors:  Noor Al Mortadi; Karem H Alzoubi; Robert Williams
Journal:  Clin Cosmet Investig Dent       Date:  2020-11-19

Review 3.  Assessment of the fit of removable partial denture fabricated by computer-aided designing/computer aided manufacturing technology.

Authors:  Khalid A O Arafa
Journal:  Saudi Med J       Date:  2018-01       Impact factor: 1.484

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.