Literature DB >> 16606391

Continuous template collection and updating for electrogram morphology discrimination in implantable cardioverter defibrillators.

Steven J Compton1, James J Merrill, Paul Dorian, Jian Cao, Duo Zhou, Jeffrey M Gillberg.   

Abstract

INTRODUCTION: Electrogram morphology analysis improves discrimination of supraventricular tachycardias (SVTs) from ventricular tachycardias (VTs) in implantable cardioverter defibrillators (ICDs), but electrogram morphology may change with lead maturation, drugs, or disease progression. We report the clinical performance of an automatic algorithm that creates and updates templates from non-paced, slow rhythm and continuously checks the quality of the template used for arrhythmia discrimination. METHODS AND
RESULTS: We studied this algorithm in 193 patients with single-chamber ICDs (Marquis VR, Medtronic Inc., Minneapolis, MN, USA). Of the 112 patients who completed 6-month follow-up, 99.1% of the patients had > or =1 automatic template created. Match scores between template and ongoing rhythm are computed using Haar Wavelets. Of the 435 automatic templates evaluated at follow-up, 423 (97.2%) had a median match score > or =70%. Intrinsic rhythm at 1 month had significantly higher match scores (P < 0.001) with automatic templates (90.3 +/- 7.0%) than with manual templates (85.7 +/- 10.9%) generated at pre-hospital discharge (PHD). The percentage of appropriately rejected SVTs was slightly higher with the automatic template (280/339 episodes) than with the manual template at PHD (272/339 episodes) while the Wavelet detection of VT was the same (218/220 episodes).
CONCLUSIONS: In patients receiving ICDs, the automatic templates were successfully created during a 6-month follow-up period, and consistently matched the patients' intrinsic rhythm at the nominal match threshold. Both early (<1 month postimplant) and late (1- to 3-month follow-up period) changes in electrogram morphology were identified, confirming the need for automatic template updating.

Entities:  

Mesh:

Year:  2006        PMID: 16606391     DOI: 10.1111/j.1540-8159.2006.00330.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Standard template of adult magnetocardiogram.

Authors:  Akihiko Kandori; Kuniomi Ogata; Tsuyoshi Miyashita; Yasushi Watanabe; Kimio Tanaka; Masahiro Murakami; Yuji Oka; Hiroshi Takaki; Syuji Hashimoto; Yuko Yamada; Kazuo Komamura; Wataru Shimizu; Shiro Kamakura; Shigeyuki Watanabe; Iwao Yamaguchi
Journal:  Ann Noninvasive Electrocardiol       Date:  2008-10       Impact factor: 1.468

Review 2.  Machine Learning in Arrhythmia and Electrophysiology.

Authors:  Natalia A Trayanova; Dan M Popescu; Julie K Shade
Journal:  Circ Res       Date:  2021-02-18       Impact factor: 17.367

3.  Influence of myopotential interference on the Wavelet discrimination algorithm in implantable cardioverter-defibrillator.

Authors:  Kazuya Mizukami; Hisashi Yokoshiki; Hirofumi Mitsuyama; Masaya Watanabe; Taro Tenma; Rui Kamada; Masayuki Takahashi; Ryo Sasaki; Motoki Maeno; Hiroyuki Tsutsui
Journal:  J Arrhythm       Date:  2016-10-01
  3 in total