Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC).

F C P du Plessis1, A Leal, S Stathakis, W Xiong, C-M Ma.   

Abstract

A study is presented that characterizes megavoltage electron beams delivered through an existing double-focused photon multi-leaf collimator (pMLC) using film measurements in a solid water phantom. Machine output stability and linearity were evaluated as well as the effect of source-to-surface distance (SSD) and field size on the penumbra for electron energies between 6 and 18 MeV over an SSD range of 60-100 cm. Penumbra variations as a function of field size, depth of measurement and the influence of the jaws were also studied. Field abutment, field flatness and target coverage for segmented beams were also addressed. The measured field size for electrons transported through the pMLC was the same as that for an x-ray beam up to SSDs of 70 cm. At larger SSD, the lower energy electron fields deviated from the projected field. Penumbra data indicated that 60 cm SSD was the most favourable treatment distance. Backprojection of P(20-80) penumbra data yielded a virtual source position located at 98.9 cm from the surface for 18 MeV electrons. For 6 MeV electrons, the virtual source position was at a distance of 82.6 cm. Penumbra values were smaller for small beam slits and reached a near-constant value for field widths larger than 5 cm. The influence of the jaws had a small effect on the penumbra. The R90 values ranged from 1.4 to 4.8 cm between 6 and 21 MeV as measured at 60 cm SSD for a 9 x 9 cm2 field. Uniformity and penumbra improvement could be demonstrated using weighted abutted fields especially useful for small segments. No detectable electron leakage through the pMLC was observed. Bremsstrahlung measurements taken at 60 cm SSD for a 9 x 9 cm2 field as shaped by the pMLC compared within 1% to bremsstrahlung measurements taken at 100 cm SSD for a 10 x 10 cm2 electron applicator field at 100 cm SSD.

Entities:  

Mesh:

Year:  2006        PMID: 16585849     DOI: 10.1088/0031-9155/51/8/011

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

1.  Monte Carlo simulations of electron beams collimated with a dual electron multileaf collimator: a feasibility study.

Authors:  S O Inyang; A C Chamberlain
Journal:  Radiol Phys Technol       Date:  2009-07-04

2.  Electron modulated arc therapy (EMAT) using photon MLC for postmastectomy chest wall treatment I: Monte Carlo-based dosimetric characterizations.

Authors:  Chaoqiong Ma; David Parsons; Mingli Chen; Steve Jiang; Qing Hou; Xuejun Gu; Weiguo Lu
Journal:  Phys Med       Date:  2019-10-10       Impact factor: 2.685

3.  Implementation and experimental validation of a robust hybrid direct aperture optimization approach for mixed-beam radiotherapy.

Authors:  Emily Heath; Silvan Mueller; Gian Guyer; Alisha Duetschler; Olgun Elicin; Daniel Aebersold; Michael K Fix; Peter Manser
Journal:  Med Phys       Date:  2021-10-14       Impact factor: 4.506

4.  Dosimetric verification of gated delivery of electron beams using a 2D ion chamber array.

Authors:  S A Yoganathan; K J Maria Das; D Gowtham Raj; Shaleen Kumar
Journal:  J Med Phys       Date:  2015 Apr-Jun

5.  Design and evaluation of electron beam energy degraders for breast boost irradiation.

Authors:  Jong In Park; Sung Whan Ha; Jung-In Kim; Hyunseok Lee; Jaegi Lee; Il Han Kim; Sung-Joon Ye
Journal:  Radiat Oncol       Date:  2016-08-31       Impact factor: 3.481

6.  Dosimetric characteristics of an electron multileaf collimator for modulated electron radiation therapy.

Authors:  Ahmed Abdel Rahman Eldib; Mohamed I ElGohary; Jiajin Fan; Lihui Jin; Jinsheng Li; Charlie Ma; Nader A Elsherbini
Journal:  J Appl Clin Med Phys       Date:  2010-04-12       Impact factor: 2.102

7.  Monte Carlo simulation for scanning technique with scattering foil free electron beam: A proof of concept study.

Authors:  Wonmo Sung; Jong In Park; Jung-In Kim; Joel Carlson; Sung-Joon Ye; Jong Min Park
Journal:  PLoS One       Date:  2017-05-11       Impact factor: 3.240

8.  Practical Dosimetry Considerations for Small MLC-Shaped Electron Fields at 60 cm SSD.

Authors:  Déte Van Eeden; Karl N Sachse; Freek C P Du Plessis
Journal:  J Biomed Phys Eng       Date:  2022-02-01

9.  Measurement and Monte Carlo simulation for energy- and intensity-modulated electron radiotherapy delivered by a computer-controlled electron multileaf collimator.

Authors:  Lihui Jin; Ahmed Eldib; Jinsheng Li; Ismail Emam; Jiajin Fan; Lu Wang; C-M Ma
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.