| Literature DB >> 16574424 |
C Michotte1, A K Pearce, M G Cox, J-J Gostely.
Abstract
The measurement model used to determine ionization chamber efficiency curves accounts from the outset for impurity corrections and beta spectrum shapes. The curves are represented by exponentials of polynomials whose coefficients are adjusted using non-linear least-squares minimization. The curves are validated by comparing with SIR key comparison reference values (KCRVs) and other published curves. The associated covariance matrix is also evaluated. Deviations from model predictions for 65Zn and 201Tl using recommended nuclear data are studied.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16574424 DOI: 10.1016/j.apradiso.2006.02.014
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513