Literature DB >> 16574424

An approach based on the SIR measurement model for determining the ionization chamber efficiency curves, and a study of 65Zn and 201Tl photon emission intensities.

C Michotte1, A K Pearce, M G Cox, J-J Gostely.   

Abstract

The measurement model used to determine ionization chamber efficiency curves accounts from the outset for impurity corrections and beta spectrum shapes. The curves are represented by exponentials of polynomials whose coefficients are adjusted using non-linear least-squares minimization. The curves are validated by comparing with SIR key comparison reference values (KCRVs) and other published curves. The associated covariance matrix is also evaluated. Deviations from model predictions for 65Zn and 201Tl using recommended nuclear data are studied.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16574424     DOI: 10.1016/j.apradiso.2006.02.014

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Revision of the NIST Standard for (223)Ra: New Measurements and Review of 2008 Data.

Authors:  B E Zimmerman; D E Bergeron; J T Cessna; R Fitzgerald; L Pibida
Journal:  J Res Natl Inst Stand Technol       Date:  2015-03-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.