The effect of axial length on photodynamic therapy.

Siamak Ansari-Shahrezaei1, Erdem Ergun, Michael Stur.   

Abstract

PURPOSE: To determine the effect of the axial length of the eye on photodynamic therapy (PDT) light fluence.
DESIGN: Laboratory investigation.
METHODS: A calibrated Gullstrand-type model eye was used for this study. The axial length of the model eye was set to different values ranging from 20 to 31 mm, and the light fluence rate for treating a spot of 4 mm was recorded and compared with a fluence rate of 600 mW/cm2 with four indirect condensing laser lenses.
RESULTS: The axial length of the eye is inversely related to light fluence. From myopia to hyperopia, the total deviation from 600 mW/cm2 was -37.7% to +53.8% (Mainster Standard; Ocular Instruments Inc, Bellevue, Washington, USA), and -48.3% to +20.5% (Mainster Wide Field; Ocular Instruments Inc), -52.3% to +16.7% (Volk Transequator; Volk Optical Inc, Mentor, Ohio, USA), and -52.3% to +15.3% (Volk PDT Laser; Volk Optical Inc).
CONCLUSIONS: Axial length of the eye has a considerable effect on PDT light fluence rate in this model. These findings may be important for optimizing PDT, particularly in eyes with high refractive errors.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16564805     DOI: 10.1016/j.ajo.2005.11.060

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  5 in total

1.  Magnification-corrected photodynamic therapy.

Authors:  Siamak Ansari-Shahrezaei; Erdem Ergun; Robert Chong; Adnan Tufail; Andreas Wedrich; Michael Stur
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2006-10-06       Impact factor: 3.117

2.  Intravitreal anti-vascular endothelial growth factor combined with half-fluence photodynamic therapy for choroidal neovascularization in chronic central serous chorioretinopathy.

Authors:  E Smretschnig; S Hagen; C Glittenberg; R Ristl; I Krebs; S Binder; S Ansari-Shahrezaei
Journal:  Eye (Lond)       Date:  2016-03-11       Impact factor: 3.775

3.  The effect of laser unit on photodynamic therapy spot size.

Authors:  Siamak Ansari-Shahrezaei; Susanne Binder; Michael Stur
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-07-27       Impact factor: 3.117

4.  The effect of photodynamic therapy on macular sensitivity in eyes with acute central serous chorioretinopathy.

Authors:  Stefan Hagen; Siamak Ansari-Shahrezaei; Eva Smretschnig; Carl Glittenberg; Ilse Krebs; Alexandra Graf; Susanne Binder
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08-29       Impact factor: 3.117

5.  Factors affecting laser power in retinal Navilas laser treatment.

Authors:  Alexandra E Hoeh; Stefanie Pollithy; Stefan Dithmar
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-08-27       Impact factor: 3.117

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.