Multi-slice computer tomography of left ventricular function with automated analysis software in comparison with conventional ventriculography.

Martine Gilard1, Pierre-Yves Pennec, Jean-Christophe Cornily, Ulric Vinsonneau, Gregoire Le Gal, Michel Nonent, Jacques Mansourati, Jacques Boschat.   

Abstract

PURPOSE: To evaluate the accuracy of left ventricular volumetric and functional parameters from multi-slice computed tomography using automated analysis software, and to correlate results with those of invasive left ventriculography.
MATERIALS AND METHODS: In 145 consecutive patients (mean age, 61 years+/-12) known or suspected to have coronary artery disease, a 16-channel multi-slice computed tomography (Philips Mx8000 IDT 16) was performed using a standard technique. Using short-axis secondary multi-slice computed tomography reformations, we determined end-diastolic and end-systolic left ventricular volumes and ejection fraction with a commercially available analysis software capable of automated contour detection. Conventional left ventriculography was performed according to standard techniques within the following 24 h. Bland-Altman analysis was performed to calculate the limits of agreement and systematic errors between multi-slice computed tomography and conventional left ventriculography.
RESULTS: As determined by computer tomography, mean end-systolic (53+/-29 mL) left ventricular volumes had an acceptable correlation with conventional ventriculography (67+/-50 mL; r=0.74; p<0.001) and mean end-diastolic (119+/-33 mL) left ventricular volumes a poor correlation with conventional ventriculography measurements (154+/-69 mL; r=0.41). Left ventricular ejection fraction (57%+/-14 versus 55%+/-14 for conventional ventriculography; r=0.79) showed a very good correlation (p<0.001). Bland-Altman analysis showed acceptable limits of agreement (+/-9.2% for ejection fraction) without systematic errors.
CONCLUSION: The use of a multi-slice computed tomography with an automatic calculation software has a good correlation with conventional ventriculography findings and could accurately assess left ventricular function, but should not be used for ventricular volumes, because of biased estimations.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16564665     DOI: 10.1016/j.ejrad.2006.02.010

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  1 in total

1.  Automatic Alignment of Myocardial Perfusion Images With Contrast-Enhanced Cardiac Computed Tomography.

Authors:  Tracy L Faber; Cesar A Santana; Marina Piccinelli; Jonathon A Nye; John R Votaw; Ernest V Garcia; Eldad Haber
Journal:  IEEE Trans Nucl Sci       Date:  2011-10       Impact factor: 1.679

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.