Literature DB >> 16564262

Prevalence and avoidance of patient-prosthesis mismatch in aortic valve replacement in small adults.

Yoshimasa Sakamoto1, Kazuhiro Hashimoto, Hiroshi Okuyama, Hiromitsu Takakura, Shinichi Ishii, Shingo Taguchi, Hiroshi Kagawa.   

Abstract

BACKGROUND: It is still controversial as to whether a small prosthesis should be inserted or a small aortic annulus should be enlarged to minimize patient-prosthesis mismatch (PPM). This retrospective study reviewed our strategy for avoiding PPM.
METHODS: Isolated or combined aortic valve replacement was performed in 181 patients, including 24 patients (13.3%) aged less than 65 years with a small aortic annulus (< or = 21 mm) who underwent enlargement of the annulus by the Manouguian (n = 18) or Nicks (n = 6) procedure. In patients aged 65 years or more, a Carpentier-Edwards Perimount pericardial (CEP) valve was implanted with few exceptions. We assessed our strategy for avoiding PPM by comparison with published normal reference values for the indexed effective orifice area.
RESULTS: A CEP valve was implanted in 53 patients, and St. Jude Medical (SJM) mechanical valves were used in 128 patients. A standard 21-mm SJM valve was only used in 4 patients and no 19-mm valves were employed. However, 19-mm CEP valves were used in 12 older patients with a small body surface area (1.43 +/- 0.14 m2). No patient receiving an SJM valve had an indexed effective orifice area of 0.85 cm2/m2 or less, and PPM developed in only 2 (3.8%) of 53 patients receiving CEP valves. Consequently, the prevalence of PPM was 1.1%. The 10-year survival rates of patients receiving CEP or SJM valves with or without annular enlargement were similar.
CONCLUSIONS: The prevalence of PPM was low in patients more than 65 years old with a relatively small body size who received bioprosthetic valves. In patients less than 65 years old with a small annulus, the method of first choice for avoiding PPM is aortic annular enlargement.

Entities:  

Mesh:

Year:  2006        PMID: 16564262     DOI: 10.1016/j.athoracsur.2005.10.013

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

Review 1.  Update on aortic valve prosthesis-patient mismatch in Japan.

Authors:  Yoshimasa Sakamoto; Kazuhiro Hashimoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-04-13

Review 2.  Beginning and development of surgery for acquired valvular heart disease in Japan.

Authors:  Kazuhiro Hashimoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2009-11-12

3.  Rupture of equine pericardial aortic-root patch after aortic valve replacement with aortic annulus enlargement: a case report.

Authors:  Akimasa Morisaki; Yasuyuki Kato; Manabu Motoki; Yosuke Takahashi; Shinsuke Nishimura; Toshihiko Shibata
Journal:  J Cardiothorac Surg       Date:  2014-06-19       Impact factor: 1.637

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.