Natalie Pomerleau-Dalcourt1, Lothar Lilge.
Abstract
Multi-sensory fluence rate probes (MSPs) yield several simultaneous measurements of photodynamic therapy (PDT) treatment light fluence from a single interstitial probe. Fluorescent sensors are embedded at desired positions along the axis of the optical fibre. A single fluorescence emission spectrum is obtained and decomposed using a partial least squares (PLS)-based analysis to yield the fluence at each sensor's location. The responsivity, linearity and possible photodegradation of each fluorophore chosen for the MSPs were evaluated using single-sensor probes. The performance of two- and three-sensor MSPs was evaluated experimentally. Individual fluorescence spectra collected from each sensor on the MSP were used to construct the training set necessary for the PLS-based analysis. The MSPs' responsivity, spatial resolution and accuracy were evaluated relative to a single scattering-tip detector. Three-fluorophore MSPs permitted three simultaneous measurements of the fluence rate gradient in a tissue-like phantom, with an average accuracy of 6.7%. No appreciable photodegradation or cross-talk was observed.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16552115 DOI: 10.1088/0031-9155/51/7/021
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609