OBJECTIVE: To compare the caries diagnostic accuracy of two software modalities used in the assessment of digital radiographs obtained with four different dental systems, and to evaluate whether the software used for image assessment influenced the mutual comparison of those four dental systems relating to their caries diagnostic accuracy.
METHODS: Under in vitro and standardized conditions 122 teeth (with 228 unrestored approximal and 99 occlusal surfaces) were radiographed in blocks of 3 test teeth and 2 non-test teeth using two storage phosphor plate systems: Digora (Soredex, Helsinki, Finland) and DenOptix (Gendex, Dentsply, Milan, Italy) and two charge coupled device (CCD)-based sensor systems: Dixi (Planmeca, Helsinki, Finland) and Sidexis (Sirona, Bensheim, Germany). The images were displayed and examined in two software modalities: their own dedicated software and a general software. Three observers examined all images for the presence of approximal enamel and dentine and occlusal dentine caries lesions using a 5-point confidence scale. The true presence of caries was validated by ground section histology. The diagnostic accuracy of the software modalities was expressed as ROC curve areas (A(z)) and differences between modalities were tested by paired t-test. Comparison of systems was analysed by post hoc t-test.
RESULTS: Results of approximal and occlusal surfaces assessed together revealed nearly identical mean A(z) with the two software modalities on images obtained with the Digora (A(z)=0.71) and DenOptix (A(z)=0.72) systems. On Dixi images the mean A(z) was 0.75 using the system's own software and 0.73 using the general software. On Sidexis images the corresponding mean A(z)s were 0.79 and 0.75. None of those differences were significant.
CONCLUSION: No significant difference in caries diagnostic accuracy was found between two software modalities used for examination of digital radiographs obtained with four different digital systems, and the software modality did not influence the mutual rank of the four systems relating to their diagnostic accuracy.

Entities:  

Mesh:

Year:  2006        PMID: 16549433     DOI: 10.1259/dmfr/50356588

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  4 in total

1.  Effect of tube potential and image receptor on the detection of natural proximal caries in primary teeth.

Authors:  Elif Sogur; B Güniz Baksı; Kaan Orhan; S Candan Paksoy; Salih Dogan; Yılmaz S Erdal; Ali Mert
Journal:  Clin Oral Investig       Date:  2010-09-14       Impact factor: 3.573

2.  Accuracy of direct digital radiography for detecting occlusal caries in primary teeth compared with conventional radiography and visual inspection: an in vitro study.

Authors:  P R Dias da Silva; M Martins Marques; W Steagall; F Medeiros Mendes; C A Lascala
Journal:  Dentomaxillofac Radiol       Date:  2010-09       Impact factor: 2.419

Review 3.  Radiographic modalities for diagnosis of caries in a historical perspective: from film to machine-intelligence supported systems.

Authors:  Ann Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2021-03-04       Impact factor: 3.525

4.  In vitro evaluation of proximal carious lesions using digital radiographic systems.

Authors:  Mayana Soares Vieira; Caroline Parente Ribeiro Nogueira; Marcos André dos Santos Silva; José Roberto de Oliveira Bauer; Etevaldo Matos Maia Filho
Journal:  ScientificWorldJournal       Date:  2015-01-27
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.