Literature DB >> 16548182

[Comparison of coronary artery bypass graft imaging between 64-slice and 16-slice spiral CT].

Zhu-hua Zhang1, Zheng-yu Jin, Ling-yan Kong, Yi-ning Wang, Lan Song, Yun Wang, Lin-hui Wang, Wen-min Zhao, Wen-bin Mou, Li-ren Zhang, Dong-jing Li, Song-Bai Lin, Shu-yang Zhang, Qi Miao.   

Abstract

OBJECTIVE: To compare the coronary artery bypass graft (CABG) imaging between 16-slice spiral CT and 64-slice spiral CT.
METHODS: Totally 27 patients with CABG received multi-slice spiral CT imaging and the results were retrospectively analyzed. Among them, 8 patients received 16-slice spiral CT scanning, 19 patients received 64-slice spiral CT scanning.
RESULTS: The evaluability rates of 64-slice spiral CT in evaluating the proximal anastomosis, bypass graft, distal anastomosis, and distal blood vessel were 100%, 100%, 90.2%, and 93.9%, respectively, while those of 16-slice spiral CT were 92.3%, 95.2%, 90.0%, and 90.0%, respectively. The patency rates of the above four aspects of 64-slice spiral CT were 66.7%, 70.0%, 71.7%, and 70.0%, respectively, while those of 16-slice spiral CT were 83.3%, 85.0%, 83.3%, and 88.9% .
CONCLUSIONS: 64-slice spiral CT is superior to 16-slice spiral CT in CABG imaging. It can be used as a non-invasive tool for the post-operative follow-up of CABG.

Entities:  

Mesh:

Year:  2006        PMID: 16548182

Source DB:  PubMed          Journal:  Zhongguo Yi Xue Ke Xue Yuan Xue Bao        ISSN: 1000-503X


  2 in total

1.  Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique.

Authors:  Oliver Klass; Martin Jeltsch; Sebastian Feuerlein; Horst Brunner; Hans-Dieter Nagel; Matthew J Walker; Hans-Juergen Brambs; Martin H K Hoffmann
Journal:  Eur Radiol       Date:  2008-11-15       Impact factor: 5.315

2.  Non-invasive assessment of coronary artery bypass graft patency using 16-slice computed tomography angiography.

Authors:  Emma S Houslay; Tristan Lawton; Anshuman Sengupta; Neal G Uren; Graham McKillop; David E Newby
Journal:  J Cardiothorac Surg       Date:  2007-06-05       Impact factor: 1.637

  2 in total