Abstract

A versatile orthogonal-coil radio frequency (RF) probe suitable for detecting the resonant frequency of miniature implantable passive sensors has been designed and tested. The probe sensitivity has been tested using printed-circuit spiral inductors of various sizes (3-15 mm) in series with discrete surface-mount capacitors designed to resonate over a range of frequencies (50-200 MHz). Close agreement between theoretical calculations and experimental results has been obtained. An equation is derived for transmit/receive (T/R) isolation that agrees with experimental measurements over the frequency range 1-500 MHz. The probe includes an additional coil to compensate for the effect of eddy currents in the human body on the probe. T/R isolation of at least 90 dB over the frequency range 1-100 MHz can be achieved when the probe is placed in close proximity to the human body.

Entities:  

Mesh:

Year:  2006        PMID: 16532781     DOI: 10.1109/TBME.2005.864496

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  1 in total

1.  The Responsivity of a Miniaturized Passive Implantable Wireless Pressure Sensor.

Authors:  Hao Jiang; Di Lan; Ken Goldman; Mozziyar Etemadi; Hamid Shahnasser; Shuvo Roy
Journal:  Proc IEEE Radio Wirel Symp       Date:  2011