Use of the Given Patency System for the screening of patients at high risk for capsule retention.

C Signorelli1, E Rondonotti, F Villa, C Abbiati, G Beccari, E C Avesani, M Vecchi, R de Franchis.   

Abstract

BACKGROUND: Capsule enteroscopy is a non-invasive diagnostic tool for the study of the small bowel. Due to the risk of capsule retention, capsule enteroscopy is contraindicated in patients with suspected small bowel strictures. The Given Patency Capsule is a disintegration time-controlled capsule developed to identify patients with strictures that may cause capsule enteroscopy retention. The presence of the patency capsule within the patient's body can be detected by a radio-frequency scanner. AIM OF THE STUDY: To evaluate safety and usefulness of the patency capsule in preventing capsule retention in patients at high risk. PATIENTS AND METHODS: Thirty-two patients were studied. Indications for patency capsule were: (A) Crohn's disease (18), (B) previous intestinal surgery (7), (C) previous obstruction (1), A+B (3), A+C (1), B+C (2). Patients were evaluated with the scanner at 72 h from ingestion.
RESULTS: At 72 h, 24 patients had already excreted the intact capsule in the stool. Of these, two experienced abdominal pain during capsule passage. In the other eight patients, the scanner detected the presence of the patency capsule. Four of them excreted the capsule intact in the stool after 72-96 h, the remaining four never found the capsule in the stool. The 26 patients who excreted the patency capsule intact without experiencing abdominal pain were deemed eligible for the capsule enteroscopy procedure, which was performed uneventfully in the 25 who agreed to undergo the examination.
CONCLUSIONS: The patency capsule is useful to identify, among patients at high risk, those who can be submitted to capsule enteroscopy without risks of capsule retention.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16527556     DOI: 10.1016/j.dld.2006.01.010

Source DB:  PubMed          Journal:  Dig Liver Dis        ISSN: 1590-8658            Impact factor:   4.088


  21 in total

1.  Diagnosis and management of mid-gastrointestinal bleeding by double-balloon endoscopy.

Authors:  Yoshikazu Hayashi; Hironori Yamamoto; Tomonori Yano; Kentaro Sugano
Journal:  Therap Adv Gastroenterol       Date:  2009-03       Impact factor: 4.409

Review 2.  Small bowel capsule endoscopy in 2007: indications, risks and limitations.

Authors:  Emanuele Rondonotti; Federica Villa; Chris J J Mulder; Maarten A J M Jacobs; Roberto de Franchis
Journal:  World J Gastroenterol       Date:  2007-12-14       Impact factor: 5.742

3.  Retained pill camera at an entero-uracho-vesical fistula site in a patient with Crohn's disease.

Authors:  Christopher G Yheulon; Daniel C Derosa; Ronald A Gagliano
Journal:  Hawaii J Med Public Health       Date:  2013-06

Review 4.  Patency and Agile capsules.

Authors:  Angel Caunedo-Alvarez; Javier Romero-Vazquez; Juan-M Herrerias-Gutierrez
Journal:  World J Gastroenterol       Date:  2008-09-14       Impact factor: 5.742

Review 5.  Imaging of inflammatory bowel disease. How?

Authors:  Melanie P Hiorns
Journal:  Pediatr Radiol       Date:  2008-06

Review 6.  Role of small bowel capsule endoscopy in the diagnosis and management of iron deficiency anemia in elderly: a comprehensive review of the current literature.

Authors:  Adnan Muhammad; Gitanjali Vidyarthi; Patrick Brady
Journal:  World J Gastroenterol       Date:  2014-07-14       Impact factor: 5.742

Review 7.  Guidelines for video capsule endoscopy: emphasis on Crohn's disease.

Authors:  Soo-Kyung Park; Byong Duk Ye; Kyeong Ok Kim; Cheol Hee Park; Wan-Sik Lee; Byung Ik Jang; Yoon Tae Jeen; Myung-Gyu Choi; Hyun Jung Kim
Journal:  Clin Endosc       Date:  2015-03-27

Review 8.  Capsule retention: prevention, diagnosis and management.

Authors:  Emanuele Rondonotti
Journal:  Ann Transl Med       Date:  2017-05

Review 9.  Role of capsule endoscopy in inflammatory bowel disease.

Authors:  Uri Kopylov; Ernest G Seidman
Journal:  World J Gastroenterol       Date:  2014-02-07       Impact factor: 5.742

10.  Safety, reliability and limitations of the given patency capsule in patients at risk of capsule retention: a 3-year technical review.

Authors:  Aymer Jonathan Postgate; David Burling; Arun Gupta; Aine Fitzpatrick; Chris Fraser
Journal:  Dig Dis Sci       Date:  2008-03-05       Impact factor: 3.199

View more