Artifactual changes in canine blood following storage, detected using the ADVIA 120 hematology analyzer.

Tommaso Furlanello1, Silvia Tasca, Marco Caldin, Erika Carli, Carlo Patron, Massimo Tranquillo, George Lubas, Laia Solano-Gallego.   

Abstract

BACKGROUND: Artifactual changes in blood may occur as a consequence of delayed analysis and may complicate interpretation of CBC data.
OBJECTIVE: The aim of this study was to characterize artifactual changes in canine blood, due to storage, using the ADVIA 120 hematology analyzer.
METHODS: Blood samples were collected into EDTA from 5 clinically healthy dogs. Within 1 hour after blood sample collection and at 12, 24, 36 and 48 hours after storage of the samples at either 4 degrees C or room temperature (approximately 24 degrees C), a CBC was done using the ADVIA 120 and multispecies software. A linear mixed model was used to statistically evaluate significant differences in values over time, compared with initial values.
RESULTS: The HCT and MCV were increased significantly after 12 hours of collection at both 4 degrees C and 24 degrees C, and continued to increase through 48 hours. The MCHC initially decreased significantly at 12-24 hours and then continued to decrease through 48 hours at both temperatures. Changes in HCT, MCV, and MCHC were greater at 24 degrees C than at 4 degrees C at all time points. A significant increase in MPV and a decrease in mean platelet component concentration were observed at all time points at 24 degrees C. Samples stored at 24 degrees C for 48 hours had significantly higher percentages of normocytic-hypochromic RBCs, and macrocytic-normochromic RBCs, and lower platelet and total WBC counts.
CONCLUSIONS: Delayed analysis of canine blood samples produces artifactual changes in CBC results, mainly in RBC morphology and platelet parameters, that are readily detected using the ADVIA 120. Refrigeration of specimens, even after 24 hours of storage at room temperature, is recommended to improve the accuracy of CBC results for canine blood samples.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16511790     DOI: 10.1111/j.1939-165x.2006.tb00087.x

Source DB:  PubMed          Journal:  Vet Clin Pathol        ISSN: 0275-6382            Impact factor:   1.180


  8 in total

1.  Cation-leak stomatocytosis in standard schnauzers does not cosegregate with coding mutations in the RhAG, SLC4A1, or GLUT1 genes associated with human disease.

Authors:  Boris E Shmukler; Alicia Rivera; David H Vandorpe; Jessica Alves; Ugo Bonfanti; Saverio Paltrinieri; Seth L Alper
Journal:  Blood Cells Mol Dis       Date:  2012-03-09       Impact factor: 3.039

2.  Platelet distribution width: a simple, practical and specific marker of activation of coagulation.

Authors:  E Vagdatli; E Gounari; E Lazaridou; E Katsibourlia; F Tsikopoulou; I Labrianou
Journal:  Hippokratia       Date:  2010-01       Impact factor: 0.471

3.  Artifactual changes in Sprague-Dawley rat hematologic parameters after storage of samples at 3 °C and 21 °C.

Authors:  Michelle C Cora; Debra King; Laura J Betz; Ralph Wilson; Greg S Travlos
Journal:  J Am Assoc Lab Anim Sci       Date:  2012       Impact factor: 1.232

4.  Rate of manual leukocyte differentials in dog, cat and horse blood samples using ADVIA 120 cytograms.

Authors:  Martina Stirn; Andreas Moritz; Natali Bauer
Journal:  BMC Vet Res       Date:  2014-06-05       Impact factor: 2.741

5.  Effects of pre-analytical handling on selected canine hematological parameters evaluated by automatic analyzer.

Authors:  Labrini Vasileiou Athanasiou; Zoe Polizopoulou; Maria Rafaela Kalafati; George Ntararas; Vasileios Kontos
Journal:  Vet Res Forum       Date:  2016-12-15       Impact factor: 1.054

6.  A large deletion in the GP9 gene in Cocker Spaniel dogs with Bernard-Soulier syndrome.

Authors:  Fabio Gentilini; Maria Elena Turba; Fiorella Giancola; Roberto Chiocchetti; Chiara Bernardini; Markéta Dajbychova; Vidhya Jagannathan; Michaela Drögemüller; Cord Drögemüller
Journal:  PLoS One       Date:  2019-09-04       Impact factor: 3.240

7.  Improving quality control for in-clinic hematology analyzers: Common myths and opportunities.

Authors:  Helen T Michael; Mary B Nabity; C Guillermo Couto; Andreas Moritz; John W Harvey; Dennis B DeNicola; Jeremy M Hammond
Journal:  Vet Clin Pathol       Date:  2022-09       Impact factor: 1.333

8.  Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats.

Authors:  Kim-Lina Charlotte Zelmer; Natali Bauer; Andreas Moritz
Journal:  J Vet Diagn Invest       Date:  2022-03-24       Impact factor: 1.279

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.