| Literature DB >> 16510322 |
Abdolhossein Zarrin1, Mahshid Foroozesh, Mehrdad Hamidi, Soleyman Mohammadi-Samani.
Abstract
Development of reliable assay methods for quantitation of interferons in dosage forms has encountered serious limitations because of the physicochemical nature of these proteins as well as the sensitivity/selectivity issues. A rapid, available, and easy-to-use reversed-phase HPLC method has been developed for quantitative analysis of interferon-alpha2b in pharmaceuticals. The reversed- phase method was based on a gradient system using a wide-pore C4 column and produced linear response in drug concentration range of 0.25-5 MIU (r = 0.9997). The average within-run and between-run variables of the method were 4.19 and 9.40%, respectively, with corresponding average accuracies of 99.48+/-4.11 and 102.83+/-9.51%. The limits of detection (LOD) and quantitation (LOQ) of the method were 0.125 and 0.25MIU/ml, respectively. The practical applicability of the method was proven throughout a post-marketing quality control program on interferon-alpha2b products (PD-feron) produced and marketed in Iran.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16510322 DOI: 10.1016/j.jchromb.2006.01.029
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205