Percutaneous valve replacement: significance of different delivery systems in vitro and in vivo.

Tim Attmann1, Georg Lutter, René Quaden, Thomas Jahnke, Kristin Rumberg, Jochen Cremer, Stefan Muller-Hulsbeck.   

Abstract

BACKGROUND AND
PURPOSE: Percutaneous heart valve replacement is an exciting growing field in cardiovascular medicine yet still with some major problems. Only sophisticated improvement of the instruments could make it a real alternative to conventional surgery. Therefore, the aim of this study was to evaluate different delivery devices for percutaneous heart valve replacement in vitro and in vivo.
METHODS: A catheter prototype designed by our group, and two commercially available devices for the delivery of esophageal stents and aortic endoprostheses, were tested. After in vitro experiments, an ovine animal model of transfemoral pulmonary valve implantation was established using biological valved self-expanding stents. Only the delivery device for aortic endografts (Medtronic, Talent, Santa Rosa, CA, USA) allowed fast in vitro procedures without material fatigue. This device was chosen for the in vivo tests.
RESULTS: Technical success was achieved in 9 of 10 animals (90%). One animal died after perforation of the ventricular wall. Orthotopic pulmonary placement was performed in 6 animals and intentional supravalvular valved stent placement in 3 animals.
CONCLUSIONS: An adequate in vitro model for this evolving field of interventional heart valve replacement is presented. Furthermore, the present study pinpoints the key characteristics that are mandatory for a delivery system in percutaneous pulmonary valve implantation. With regard to the delivery device's ductility observed during this "venous" study, an approach to transfemoral aortic valve implantation seems feasible.

Entities:  

Mesh:

Year:  2006        PMID: 16502169     DOI: 10.1007/s00270-005-0244-4

Source DB:  PubMed          Journal:  Cardiovasc Intervent Radiol        ISSN: 0174-1551            Impact factor:   2.740


  1 in total

1.  Evaluation of Hybrid Surgical Access Approaches for Pulmonary Valve Implantation in an Acute Swine Model.

Authors:  Ruth Thalmann; Elena M Merkel; Bassil Akra; Rene Bombien; Rainer G Kozlik-Feldmann; Christoph Schmitz
Journal:  Comp Med       Date:  2019-06-20       Impact factor: 0.982

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.