| Literature DB >> 16495621 |
T S Toellner1, A Alatas, A Said, D Shu, W Sturhahn, J Zhao.
Abstract
The design and performance results for a cryogenically stabilized high-resolution monochromator for 23.880 keV (lambda = 52 pm) X-rays are presented. The four-crystal-reflection monochromator is suitable for nuclear resonant scattering measurements from 119Sn compounds using synchrotron radiation. The design includes a low-vibration cryostat that maintains two of the four crystal reflections at a temperature where the coefficient for thermal expansion of the crystalline material (silicon) vanishes. Test results demonstrate a 1.3 meV bandwidth with negligible broadening due to vibrations and a spectral efficiency of 37% when used with an undulator source.Entities:
Year: 2006 PMID: 16495621 DOI: 10.1107/S0909049505042664
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616