Literature DB >> 16493679

Ultrastructural evaluation of multiple pass low energy versus single pass high energy radio-frequency treatment.

David Kist1, A Jay Burns, Roth Sanner, Jeff Counters, Brian Zelickson.   

Abstract

BACKGROUND AND
OBJECTIVE: The radio-frequency (RF) device is a system capable of volumetric heating of the mid to deep dermis and selective heating of the fibrous septa strands and fascia layer. Clinically, these effects promote dermal collagen production, and tightening of these deep subcutaneous structures. A new technique of using multiple low energy passes has been described which results in lower patient discomfort and fewer side effects. This technique has also been anecdotally described as giving more reproducible and reliable clinical results of tissue tightening and contouring. This study will compare ultrastructural changes in collagen between a single pass high energy versus up to five passes of a multiple pass lower energy treatment. STUDY DESIGN/
MATERIALS AND METHODS: Three subjects were consented and treated in the preauricular region with the RF device using single or multiple passes (three or five) in the same 1.5 cm(2) treatment area with a slight delay between passes to allow tissue cooling. Biopsies from each treatment region and a control biopsy were taken immediately, 24 hours or 6 months post treatment for electron microscopic examination of the 0-1 mm and 1-2 mm levels. Sections of tissue 1 mm x 1 mm x 80 nm were examined with an RCA EMU-4 Transmission Electron Microscope. Twenty sections from 6 blocks from each 1 mm depth were examined by 2 blinded observers. The morphology and degree of collagen change in relation to area examined was compared to the control tissue, and estimated using a quantitative scale.
RESULTS: Ultrastructural examination of tissue showed that an increased amount of collagen fibril changes with increasing passes at energies of 97 J (three passes) and 122 J (five passes), respectively. The changes seen after five multiple passes were similar to those detected after much more painful single pass high-energy treatments.
CONCLUSIONS: This ultrastructural study shows changes in collagen fibril morphology with an increased effect demonstrated at greater depths of the skin with multiple low-fluence passes and at lesser depths with single pass higher fluence settings. Findings suggest that similar collagen fibril alteration can occur with multiple pass low-energy treatments and single pulse high-energy treatments. The lower fluence multiple pass approach is associated with less patient discomfort, less side effects, and more consistent clinical results. Copyright 2005 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16493679     DOI: 10.1002/lsm.20303

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  6 in total

1.  Electro-optical Synergy Technique: A New and Effective Nonablative Approach to Skin Aging.

Authors:  Moetaz El-Domyati; Tarek S El-Ammawi; Walid Medhat; Osama Moawad; My G Mahoney; Jouni Uitto
Journal:  J Clin Aesthet Dermatol       Date:  2010-12

2.  Short-term nanostructural effects of high radiofrequency treatment on the skin tissues of rabbits.

Authors:  Samjin Choi; Youjin Cheong; Jae-Ho Shin; Hui-Jae Lee; Gi-Ja Lee; Seok Keun Choi; Kyung-Hyun Jin; Hun-Kuk Park
Journal:  Lasers Med Sci       Date:  2011-10-30       Impact factor: 3.161

3.  Evaluating Clinical Use of a Ceramide-dominant, Physiologic Lipid-based Topical Emulsion for Atopic Dermatitis.

Authors:  Leon H Kircik; James Q Del Rosso; Daniel Aversa
Journal:  J Clin Aesthet Dermatol       Date:  2011-03

4.  Efficacy and Safety of Monopolar Radiofrequency for Treatment of Lower Facial Laxity in Asians.

Authors:  Rungsima Wanitphakdeedecha; Yuri Yogya; Chadakan Yan; Phumithep Phumariyapong; Yanisorn Nanchaipruek; Panwadee Thongjaroensirikul; Pitchaya Maneeprasopchoke; Thanya Techapichetvanich; Sasima Eimpunth; Woraphong Manuskiatti
Journal:  Dermatol Ther (Heidelb)       Date:  2022-09-27

5.  Radiofrequency treatment induces fibroblast growth factor 2 expression and subsequently promotes neocollagenesis and neoangiogenesis in the skin tissue.

Authors:  Patrícia Froes Meyer; Patrícia de Oliveira; Flávia K B A Silva; Ana C S da Costa; Caline R A Pereira; Sebastian Casenave; Rodrigo Marcel Valentim Silva; Luis Gonzaga Araújo-Neto; Sebastião David Santos-Filho; Eric Aizamaque; Hennes Gentil Araújo; Mario Bernardo-Filho; Maria Goretti Freire Carvalho; Ciro Dantas Soares
Journal:  Lasers Med Sci       Date:  2017-05-31       Impact factor: 3.161

6.  Objective assessment of skin tightening in Asians using a water-filtered near-infrared (1,000-1,800 nm) device with contact-cooling and freezer-stored gel.

Authors:  Yohei Tanaka; Yuichiro Tsunemi; Makoto Kawashima; Naoto Tatewaki; Hiroshi Nishida
Journal:  Clin Cosmet Investig Dermatol       Date:  2013-06-26