Abstract

AIM: We recently introduced an experimental surface detection system based on a conventional dental ultrasonic scaler. This device automatically discriminates cementum and dental calculus, which is the prerequisite for complete and thorough calculus removal. In the present study, the detection limits of this device were tested in vitro.
MATERIAL AND METHODS: From 50 extracted teeth, subgingival calculus was gradually removed using a Gracey curette. During this stepwise procedure, detection properties of the surface detection system were continuously monitored and systematically verified until the system stopped discriminating calculus from the root surface. By measuring the diameter, circumference and area of the smallest, yet recognizable deposit, and of the no longer recognizable deposit, the cut-off point of the discriminative capability of the detection device was determined.
RESULTS: The cut-off points for the correct classification of residual deposits averaged on a diameter of 219 microm, an area of 21,600 microm2, and a circumference of 748 microm. This means a sensitivity of 73% and a specificity of 80% in this critical area.
CONCLUSIONS: This calculus detection system was able to detect small deposits. In clinical practice, this device may support dentists in deciding whether to stop or to continue the debridement.

Entities:  

Mesh:

Year:  2006        PMID: 16489945     DOI: 10.1111/j.1600-051X.2006.00882.x

Source DB:  PubMed          Journal:  J Clin Periodontol        ISSN: 0303-6979            Impact factor:   8.728


  3 in total

1.  Reliability of recordings of subgingival calculus detected using an ultrasonic device.

Authors:  Priscila Corraini; Rodrigo López
Journal:  Clin Oral Investig       Date:  2014-07-24       Impact factor: 3.573

Review 2.  Dental optical coherence tomography.

Authors:  Yao-Sheng Hsieh; Yi-Ching Ho; Shyh-Yuan Lee; Ching-Cheng Chuang; Jui-che Tsai; Kun-Feng Lin; Chia-Wei Sun
Journal:  Sensors (Basel)       Date:  2013-07-12       Impact factor: 3.576

Review 3.  Calculus detection technologies: where do we stand now?

Authors:  V Archana
Journal:  J Med Life       Date:  2014
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.