Literature DB >> 16489568

Percutaneous closure of postoperative ventricular septal defects with the Amplatzer device.

Mark A Walsh1, David M Coleman, Paul Oslizlok, Kevin P Walsh.   

Abstract

The objective of this study was to look at the procedure, the results, and the follow-up of patients who underwent percutaneous closure of a residual ventricular septal defect (VSD) following a surgical closure using the Amplatzer VSD device. Four patients had an original diagnosis of tetralogy of Fallot, two patients had a patch leak following a surgical repair of a VSD, and three patients had a VSD not repaired at the time of surgery. All patients fulfilled the currently accepted surgical criterion for reoperation (Qp/Qs>1.5). The mean Qp/Qs was 1.8+/-0.3 (1.5-2.3). Four patients underwent VSD closure using an Amplatzer perimembranous VSD device and in five patients an Amplatzer muscular VSD device was implanted. We performed percutaneous closure in nine patients. The size of the residual shunt ranged from 6 to 14 mm and the size of device used ranged from 8 to 16 mm. The arteriovenous loop needed to be recreated in two patients because of failure to advance the delivery sheath. There was complete closure of the defect in six cases, and a small residual shunt remained in three cases. Percutaneous closure of postoperative VSDs appears to be an effective way to resolve a hemodynamically significant residual shunt. There were no difficulties encountered with implantation of the devices. These promising short-term results need reinforcement with additional long-term data. Copyright (c) 2006 Wiley-Liss, Inc.

Entities:  

Mesh:

Year:  2006        PMID: 16489568     DOI: 10.1002/ccd.20626

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  4 in total

1.  Transcatheter Closure of Residual and Iatrogenic Ventricular Septal Defects: Tertiary Center Experience and Outcome.

Authors:  Fatma A Taha; Fareed Alnozha; Osama Amoudi; Mansour Almutairi; Reda Abuelatta
Journal:  Pediatr Cardiol       Date:  2021-09-08       Impact factor: 1.655

2.  Transcatheter closure of ventricular septal defects with nitinol wire occluders of type patent ductus arteriosus.

Authors:  Arkadiusz Wierzyk; Małgorzata Szkutnik; Roland Fiszer; Paweł Banaszak; Szymon Pawlak; Jacek Białkowski
Journal:  Postepy Kardiol Interwencyjnej       Date:  2014-03-23       Impact factor: 1.426

3.  Transcatheter Closure Versus Repeat Surgery for the Treatment of Postoperative Left-to-Right Shunts: A Single Center 15-Year Experience.

Authors:  Xinghua Gu; Qiuwang Zhang; Hourong Sun; Jianchun Fei; Xiquan Zhang; Michael J Kutryk
Journal:  Cardiol Res       Date:  2017-12-22

4.  The experience of transcatheter closure of postoperative ventricular septal defect after total correction.

Authors:  N'goran Yves N'da Kouakou; Jinyoung Song; June Huh; I-Seok Kang
Journal:  J Cardiothorac Surg       Date:  2019-06-11       Impact factor: 1.637

  4 in total