Using a human visual system model to optimize soft-copy mammography display: influence of veiling glare.

Elizabeth A Krupinski1, Jeffrey Lubin, Hans Roehrig, Jeffrey Johnson, John Nafziger.   

Abstract

RATIONALE AND
OBJECTIVES: This project evaluated human observer performance and that of a human visual system model (JNDmetrix) to assess whether the veiling glare of a digital display influences observer performance during soft-copy interpretation of mammographic images for the detection of masses.
MATERIALS AND METHODS: A set of 160 mammographic images, half containing a single mass, was processed to simulate four levels of veiling glare: none, comparable to a medical grade monochrome curved-screen cathode ray tube (CRT) display, double that of the CRT and quadruple that of the CRT. The images were shown to six observers in a randomized presentation order on a liquid crystal display (LCD) that had essentially no veiling glare. The images were also analyzed using the JNDmetrix human visual system model.
RESULTS: Observer performance as measured using receiver operating characteristic techniques declined with increasing veiling glare (F = 6.884, P = .0035), with quadruple veiling glare yielding significantly lower performance than the lower veiling glare levels. The JNDmetrix model did not show a large reduction in performance as a function of veiling glare, and correlation with the human observer data was modest (0.588).
CONCLUSIONS: Soft-copy display veiling glare can influence observer performance, but only at extreme levels. The impact of veiling glare on performance may be more pronounced for less experienced readers.

Entities:  

Mesh:

Year:  2006        PMID: 16488840     DOI: 10.1016/j.acra.2005.11.006

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  An Evaluation of Performance Characteristics of Primary Display Devices.

Authors:  Ernest U Ekpo; Mark F McEntee
Journal:  J Digit Imaging       Date:  2016-04       Impact factor: 4.056

2.  Thousand shades of gray - The role of imaging display in diagnosis of occult scaphoid fractures - A pilot study.

Authors:  Viktor Feldman; Ran Atzmon; Jeremy Dubin; Orit Bein; Ezequiel Palmanovich; Nissim Ohana; Uri Farkash
Journal:  J Orthop       Date:  2022-09-28

3.  Exploration of analysis methods for diagnostic imaging tests: problems with ROC AUC and confidence scores in CT colonography.

Authors:  Susan Mallett; Steve Halligan; Gary S Collins; Doug G Altman
Journal:  PLoS One       Date:  2014-10-29       Impact factor: 3.240