Two-dimensional echocardiography for the diagnosis of interventricular septum perforation by a temporary pacing catheter.

Kuan-Hung Yeh1, Chi-Wen Cheng, Li-Tang Kuo, Kuo-Chun Hung.   

Abstract

This case study describes a patient with complete heart block inadvertently paced from the left ventricular posterior wall due to perforation of interventricular septum by a temporary pacing catheter. This is a rare but potentially fatal complication of pacing. The frontal chest radiograph neither identified abnormalities nor could determine the exact site of the catheter tip. The electrocardiogram revealed a right bundle branch block pattern. Echocardiography was performed immediately at bedside and the diagnosis was made. The temporary pacing catheter was removed without complications and, the next day, the patient underwent permanent pacemaker implantation. Given its noninvasiveness and availability, echocardiography is a highly effective means of assessing pacemaker complications such as catheter perforation.

Entities:  

Mesh:

Year:  2006        PMID: 16479183     DOI: 10.1097/00000441-200602000-00017

Source DB:  PubMed          Journal:  Am J Med Sci        ISSN: 0002-9629            Impact factor:   2.378


  1 in total

1.  Asymptomatic Right Ventricular Perforation by a Temporary Transvenous Pacing Lead in an Infant.

Authors:  Hsu-Ting Cheng; Yun-Ching Fu; Chung-Chi Wang; Sheng-Ching Chan; Ming-Chih Lin; Sheng-Ling Jan
Journal:  Acta Cardiol Sin       Date:  2013-07       Impact factor: 2.672