| Literature DB >> 16472814 |
Anand Sistla1, Wei-Lien Yang, Narmada Shenoy.
Abstract
SU5416 shows light-induced reversible geometric isomerism. A simple, reliable, isocratic HPLC method using an UV-vis detector at lambda(425nm) was developed. The method provides efficient (R(S)=3.5) analysis of the two isomers with retention of the isomeric integrity. Additionally, the method has linearity over a wide range (50-1000microg/mL, r(2)=0.99), is accurate (99-102%, RSD <4%), and reproducible (RSD <0.8%). The method was used for analyzing pharmaceutical samples and understanding the kinetics of SU5416 isomers in methanol. In addition, this method can be used for quantifying the non-isolatable E-isomer.Entities:
Mesh:
Substances:
Year: 2006 PMID: 16472814 DOI: 10.1016/j.chroma.2006.01.079