High-performance liquid chromatographic method for determination of reversible isomers of SU5416.

Anand Sistla1, Wei-Lien Yang, Narmada Shenoy.   

Abstract

SU5416 shows light-induced reversible geometric isomerism. A simple, reliable, isocratic HPLC method using an UV-vis detector at lambda(425nm) was developed. The method provides efficient (R(S)=3.5) analysis of the two isomers with retention of the isomeric integrity. Additionally, the method has linearity over a wide range (50-1000microg/mL, r(2)=0.99), is accurate (99-102%, RSD <4%), and reproducible (RSD <0.8%). The method was used for analyzing pharmaceutical samples and understanding the kinetics of SU5416 isomers in methanol. In addition, this method can be used for quantifying the non-isolatable E-isomer.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16472814     DOI: 10.1016/j.chroma.2006.01.079

Source DB:  PubMed          Journal:  J Chromatogr A        ISSN: 0021-9673            Impact factor:   4.759


  1 in total

1.  Quantification of sunitinib in mouse plasma, brain tumor and normal brain using liquid chromatography-electrospray ionization-tandem mass spectrometry and pharmacokinetic application.

Authors:  Qingyu Zhou; James M Gallo
Journal:  J Pharm Biomed Anal       Date:  2009-10-12       Impact factor: 3.935