Bipolar and multipolar radio frequency ablation with resistance controlled power output: standardized ex vivo kidney tissue evaluation.

Axel Häcker1, Stefan Vallo, Christel Weiss, Rainer Grobholz, Thomas Stein, Thomas Knoll, Maurice Stephan Michel.   

Abstract

PURPOSE: We investigated a newly developed bipolar and multipolar RF ablation system with an internally cooled electrode and resistance controlled power output in a standardized model of perfused ex vivo kidney tissue.
MATERIALS AND METHODS: RF energy was applied at different power levels (20, 30 and 60 W) for 1, 3, 5 and 9 minutes. Each treatment parameter was repeated 5 times. For the 20/30 W levels a bipolar electrode with an active conducting part of 20/30 mm was selected. At 60 W 2 bipolar electrodes with an active conducting part (30 mm each) were connected. Lesion volumes and shapes were calculated by measuring the maximum vertical, long axis and short axis diameters of the macroscopic lesion.
RESULTS: Lesion volume increased significantly with the treatment time and generator power applied (p < 0.0001). Lesion size in multipolar ablated zones was larger than that in bipolar ablated zones. A reliable dose-effect relationship existed between the generator power/applied treatment time and ablated tissue lesion size. All lesions were elliptical.
CONCLUSIONS: Bipolar and multipolar RF ablation with an internally cooled electrode and tissue resistance control represent an interesting advance in RF technology. The development of lesion size and volume is predictable, while a uniform lesion shape can be achieved in perfused ex vivo kidney tissue. Further in vivo trials are required to test whether complete and reliable tumor tissue ablation is possible with this system.

Entities:  

Mesh:

Year:  2006        PMID: 16469637     DOI: 10.1016/S0022-5347(05)00316-2

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  2 in total

1.  Image-based 3D modeling and validation of radiofrequency interstitial tumor ablation using a tissue-mimicking breast phantom.

Authors:  Zhigang Wang; Isshaa Aarya; Mariana Gueorguieva; Dun Liu; Hongyan Luo; Luigi Manfredi; Lijun Wang; Donald McLean; Stuart Coleman; Stuart Brown; Alfred Cuschieri
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-06-12       Impact factor: 2.924

2.  Safety and feasibility of radiofrequency ablation using bipolar electrodes for aldosterone-producing adenoma: a multicentric prospective clinical study.

Authors:  Sota Oguro; Ryo Morimoto; Kazumasa Seiji; Hideki Ota; Tomo Kinoshita; Masahiro Kawabata; Yoshikiyo Ono; Kei Omata; Yuta Tezuka; Fumitoshi Satoh; Sadayoshi Ito; Nobukazu Moriya; Seishi Matsui; Tetsuo Nishikawa; Masao Omura; Kazuki Nakai; Seishi Nakatsuka; Isao Kurihara; Kazutoshi Miyashita; Wataru Koda; Tetsuya Minami; Yoshiyu Takeda; Mitsuhiro Kometani; Yutaka Oki; Toshihiro Oishi; Takasuke Ushio; Satoshi Goshima; Kei Takase
Journal:  Sci Rep       Date:  2022-08-18       Impact factor: 4.996

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.