Variation of depth of cure and intensity with distance using LED curing lights.

Krishna Aravamudhan1, Duane Rakowski, P L Fan.   

Abstract

OBJECTIVES: The purpose of the study was to determine the correlation between intensity of light-emitting diode (LED) and tungsten-halogen light sources, and depth of cure of a resin composite at different distances.
METHODS: Four LED curing lights (Flashlite 1001, Freelight 2, Smartlite IQ and Ultralume 5) and one tungsten halogen (Optilux 501, with 8 and 11 mm tips) were evaluated. Intensity was measured according a modified ISO Standard 10650 at distances of 0, 2, 4, 6, 8, 10 mm between the light tip and detector. Depth of cure (DOC) of TPH Spectrum shade A2 was measured according to the international standard ISO 4049 at the same distances.
RESULTS: For all lights, intensity decreased as distance increased. The authors documented a logarithmic correlation between intensity and distance for all lights except the Smartlite IQ, Ultralume 5 and the Optilux 501 with the 11 mm tip, which showed a linear relationship between intensity and distance. All lights demonstrated a logarithmic correlation between intensity and DOC, and a linear correlation between DOC and distance. Smartlite IQ and Optilux 501 (11 mm tip) also had the least reduction in intensity and DOC at 10 mm. SIGNIFICANCE: Clinicians often an experience difficulty placing the light tip close to the resin surface when curing resin composites. While both intensity and DOC decrease with increasing distance, the relationship between these factors and distance may not be similar for all lights and may depend on the characteristics of individual lights.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16469373     DOI: 10.1016/j.dental.2005.11.031

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  7 in total

1.  The effect of irradiation distance on microhardness of resin composites cured with different light curing units.

Authors:  Isil Cekic-Nagas; Ferhan Egilmez; Gulfem Ergun
Journal:  Eur J Dent       Date:  2010-10

2.  Evaluation of Vickers hardness and depth of cure of six composite resins photo-activated with different polymerization modes.

Authors:  C Poggio; M Lombardini; S Gaviati; M Chiesa
Journal:  J Conserv Dent       Date:  2012-07

3.  Comparative study of surface microhardness of methacrylate-based composite resins polymerized with light-emitting diodes and halogen.

Authors:  Camila Sabatini
Journal:  Eur J Dent       Date:  2013-07

4.  Composite resin color stability: influence of light sources and immersion media.

Authors:  Patricia Aleixo dos Santos Domingos; Patrícia Petromilli Nordi Sasso Garcia; Ana Luisa Botta Martins de Oliveira; Regina Guenka Palma-Dibb
Journal:  J Appl Oral Sci       Date:  2011 May-Jun       Impact factor: 2.698

5.  Comparative depth of cure among two light-cured core build-up composites by surface vickers hardness.

Authors:  Mr Rouhollahi; M Mohammadibasir; Sh Talim
Journal:  J Dent (Tehran)       Date:  2012-09-30

6.  Influence of polymerization time and depth of cure of resin composites determined by Vickers hardness.

Authors:  Marco Lombardini; Marco Chiesa; Andrea Scribante; Marco Colombo; Claudio Poggio
Journal:  Dent Res J (Isfahan)       Date:  2012-11

7.  LED Curing Lights and Temperature Changes in Different Tooth Sites.

Authors:  E Armellin; G Bovesecchi; P Coppa; G Pasquantonio; L Cerroni
Journal:  Biomed Res Int       Date:  2016-04-18       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.