Literature DB >> 16468554

Intraoperative management of nonfunctioning Ahmed glaucoma valve implant.

Pete Setabutr1, Nicholas P F Bell, Robert M Feldman.   

Abstract

A 62-year-old woman with neovascular glaucoma underwent filtering surgery with the Ahmed glaucoma valve (New World Medical, Inc., Rancho Cucamonga, CA). Intraoperative hypotony secondary to a defective valve was noted and managed by excising the valve mechanism. The implant was then treated as if it was non-valved by placement of an obturator suture. The postoperative course was unremarkable for a non-valved implant. This case demonstrates how intraoperative hypotony secondary to overfiltration of an Ahmed glaucoma valve implant can be successfully managed by converting to a non-valved implant.

Entities:  

Mesh:

Year:  2006        PMID: 16468554

Source DB:  PubMed          Journal:  Ophthalmic Surg Lasers Imaging        ISSN: 1542-8877


  1 in total

1.  MR imaging features of giant reservoir formation in the orbit: an unusual complication of Ahmed glaucoma valve implantation.

Authors:  T Y Jeon; H-J Kim; S T Kim; T-Y Chung; C Kee
Journal:  AJNR Am J Neuroradiol       Date:  2007-09       Impact factor: 3.825