Abstract
An assessment is provided of protection factors afforded for retinal thermal hazard and blue light photochemical hazard for a range of filters used with intense pulsed light sources (IPLs). A characteristic IPL spectrum based on black body radiation at 5000 K with a low cut filter at 515 nm was identified as suitable for such estimations. Specific filters assessed included types with idealized transmission properties and also a range of types whose transmission characteristics were measured by means of a Bentham DMc150 spectroradiometer. Predicted behaviour based on these spectra is outlined which describes both the effectiveness of protection and the level of luminous transmittance afforded. The analysis showed it was possible to describe a figure of merit for a particular filter material relating the degree of protection provided and corresponding value of luminous transmittance. This consideration is important for providing users of IPL equipment with safety eyewear with adequate level of visual transmittance.Entities: