Literature DB >> 16463985

Cardiac resynchronization therapy with sequential biventricular pacing: impact of echocardiography guided VV delay optimization on acute results.

Leonor Parreira1, José Ferreira Santos, João Madeira, Lígia Mendes, Filipe Seixo, Filomena Caetano, Cláudia Lopes, José Venãncio, Arminda Mateus, J Lopes Inês, Miguel Mendes.   

Abstract

BACKGROUND: Cardiac resynchronization therapy (CRT) improves left ventricular synchrony as evaluated by tissue Doppler imaging (TDI), leading to improved left ventricular performance and reverse remodeling. New CRT devices enable programming of left and right VV delay. The aim of this study was to determine whether sequential biventricular (BiV) pacing by echo-guided programming of VV delay would enhance the response to CRT.
METHODS: 15 consecutive patients with severe heart failure and left bundle branch block underwent CRT by BiV device implantation. They were studied with conventional and TDI echo the day before implantation. Left ventricular ejection fraction (LVEF) was determined, and the electromechanical delay (QS), defined as the time interval from the beginning of the QRS to the S wave in pulsed TDI, was assessed in each of the four left ventricular basal segments. The dyssynchrony index was calculated as the difference between the longest and shortest electromechanical delay (QS(max-min)). The parameters were re-evaluated the day after implantation during simultaneous BiV pacing and with seven different VV delays. The optimal VV delay was determined by finding the VV interval corresponding to the maximum aortic velocity time interval (VTI).
RESULTS: QS(max-min) decreased from 85.3 +/- 27.0 msec to 46.7 +/- 23.0 msec (p = 0.0002), LVEF increased from 21.7 +/- 7.3% to 30.0 +/- 7.7% (p = 0.0001) and aortic VTI increased from 12.7 +/- 3.6 cm to 15.2 +/- 4.0 cm (p < 0.0001), with simultaneous BiV pacing. The VV intervals were programmed as follows: LV pre-excitation by 10 msec in five patients, 20 msec in three, 30 msec in two, and 40 msec in three; and RV pre-excitation by 10 msec in one and by 20 msec in one. The maximal aortic VTI obtained with VV delay programming increased from 15.2 +/- 4.0 cm to 17.7 +/- 4.0 cm (p = 0.0005). During optimized sequential BiV pacing, QS(max-min) further decreased from 46.7 +/- 23.0 msec to 30.6 +/- 21.0 msec (p = 0.02) and LVEF further increased from 30.0 +/- 7.7% to 35.0 +/- 7.7% (p = 0.0003).
CONCLUSIONS: Sequential BiV pacing with VV delay optimized by evaluation of aortic VTI enhanced the response to CRT with additional improvements in left ventricular synchrony and left ventricular function compared to simultaneous CRT.

Entities:  

Mesh:

Year:  2005        PMID: 16463985

Source DB:  PubMed          Journal:  Rev Port Cardiol        ISSN: 0870-2551            Impact factor:   1.374


  4 in total

Review 1.  Optimal use of echocardiography in cardiac resynchronisation therapy.

Authors:  Gabe B Bleeker; Cheuk-Man Yu; Petros Nihoyannopoulos; Johan de Sutter; Nico Van de Veire; Eduard R Holman; Martin J Schalij; Ernst E van der Wall; Jeroen J Bax
Journal:  Heart       Date:  2007-11       Impact factor: 5.994

2.  Lack of clinical predictors of optimal V-V delay in patients with cardiac resynchronization devices.

Authors:  Avi Fischer; Riple Hansalia; Samantha Buckley; Robin Goldberg; Martin Goldman; Paul Muntner; Davendra Mehta; W Lane Duvall
Journal:  J Interv Card Electrophysiol       Date:  2009-01-16       Impact factor: 1.900

3.  Acute beneficial hemodynamic effects of a novel 3D-echocardiographic optimization protocol in cardiac resynchronization therapy.

Authors:  Carolin Sonne; Lorenz Bott-Flügel; Simon Hauck; Hasema Lesevic; Petra Barthel; Fabian Michalk; Katharina Hoppe; Jörg Hausleiter; Albert Schömig; Christof Kolb
Journal:  PLoS One       Date:  2012-02-03       Impact factor: 3.240

4.  Reliability of a novel intracardiac electrogram method for AV And VV delay optimization and comparability to echocardiography procedure for determining optimal conduction delays in CRT patients.

Authors:  N Reinsch; C Buhr; S Huptas; T Buck; T Konorza; H Wieneke; R Erbel
Journal:  Indian Pacing Electrophysiol J       Date:  2009-03-15
  4 in total