Literature DB >> 16463984

Performance of the Colson MAM BP 3AA1-2 automatic blood pressure monitor according to the European Society of Hypertension validation protocol.

Telmo Pereira1, João Maldonado.   

Abstract

OBJECTIVE: To evaluate the performance of the Colson MAM BP 3AA1-2 oscillometric automatic blood pressure monitor according to the validation protocol of the European Society of Hypertension, testing its suitability for self-measurement of blood pressure. The performance of the device was assessed in relation to various clinical variables, including age, gender, body mass index, arm circumference and arterial stiffness.
METHOD: 33 subjects (15 men and 18 women), with a mean age of 47 +/- 10 years, were studied according to the procedures laid down in the European Society of Hypertension validation protocol. Sequential same-arm blood pressure measurements were made, alternating between a mercury standard and the automatic device. The differences among the test-control measurements were assessed and divided into categorization zones of 5, 10 and 15 mmHg discrepancy. Aortic pulse wave velocity was assessed in all subjects with a Complior device (Colson, Paris).
RESULTS: The Colson MAM BP 3AA1-2 passed all three phases of the protocol for both systolic and diastolic blood pressure. The mean differences between the test and control measurements were -1.0 +/- 5.0 mmHg for systolic blood pressure and -1.1 +/- 4.1 mmHg for diastolic blood pressure. Both standard deviations are well below the 8 mmHg limit proposed by the Association for the Advancement of Medical Instrumentation. The predictive value of various clinical variables for the discrepancies was assessed by a regression model analysis, with no variable being found that independently undermined the performance of the monitor. In another regression analysis, we found a similar relation between test and control blood pressures and aortic pulse wave velocity, a widely recognized and validated index of target organ damage.
CONCLUSIONS: These data show that the Colson MAM BP 3AA1-2 satisfies the quality requirements proposed by the European Society of Hypertension, demonstrating its suitability for inclusion in integrated programs of clinical surveillance based on self-measurement of blood pressure. The uniformity of its performance over a wide spectrum of clinical characteristics and the relation found with pulse wave velocity further reinforce its clinical validity.

Entities:  

Mesh:

Year:  2005        PMID: 16463984

Source DB:  PubMed          Journal:  Rev Port Cardiol        ISSN: 0870-2551            Impact factor:   1.374


  3 in total

1.  Hypertension and longevity: role of genetic polymorphisms in renin-angiotensin-aldosterone system and endothelial nitric oxide synthase.

Authors:  Alda Pereira da Silva; Andreia Matos; Laura Aguiar; Nuno Ramos-Marques; Ricardo Ribeiro; Ângela Gil; João Gorjão-Clara; Manuel Bicho
Journal:  Mol Cell Biochem       Date:  2018-11-20       Impact factor: 3.396

2.  Central arterial hemodynamic effects of dark chocolate ingestion in young healthy people: a randomized and controlled trial.

Authors:  T Pereira; J Maldonado; M Laranjeiro; R Coutinho; E Cardoso; I Andrade; J Conde
Journal:  Cardiol Res Pract       Date:  2014-05-27       Impact factor: 1.866

3.  Impact on Longevity of Genetic Cardiovascular Risk and Lifestyle including Red Meat Consumption.

Authors:  Alda Pereira da Silva; Maria do Céu Costa; Laura Aguiar; Andreia Matos; Ângela Gil; J Gorjão-Clara; Jorge Polónia; Manuel Bicho
Journal:  Oxid Med Cell Longev       Date:  2020-06-30       Impact factor: 6.543

  3 in total