Literature DB >> 16457915

Arthroscopic evaluation of cartilage degeneration using indentation testing--influence of indenter geometry.

L P Li1, W Herzog.   

Abstract

BACKGROUND: It has been suggested that the early onset of cartilage degeneration might be detected with a handheld indentation probe during knee arthroscopy, prior to any visible change on the articular surface. Collagen degradation has been considered as the first sign of cartilage degeneration. Therefore, it is important to consider the collagen network as a distinct constituent in the study of arthroscopic evaluation of cartilage degeneration.
METHODS: The tip of an arthroscopic probe (indenter) was modeled as rigid and in contact with a cartilage/bone disk of sufficiently large radius to simulate an indentation in a joint. A fibril-reinforced model of cartilage, including streaming potentials and distinct constitutive laws for the proteoglycan matrix and collagen network, was used to determine the contact mechanics of indenter and cartilage. The finite element package ABAQUS was employed to obtain numerical solutions.
FINDINGS: A spherical indenter produces a relatively uniform deformation in cartilage, but can easily slide on the articular surface. In contrast, a cylindrical indenter produces great deformation gradients for quick compression rates, but does not slide as easily on the articular surface as the spherical indenter. Small porous and large solid indenters should be used to evaluate the properties of the proteoglycan matrix and collagen network, respectively, in order to minimize or maximize the fluid pressure in the corresponding case. When the collagen network is substantially degraded, the gradients of fluid pressure and deformation are greatly reduced regardless of indenter geometry.
INTERPRETATION: The indenter geometry including its porosity is important to the material safety of articular cartilage in indentation and precise evaluation of cartilage degeneration.

Entities:  

Mesh:

Year:  2006        PMID: 16457915     DOI: 10.1016/j.clinbiomech.2005.12.010

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  8 in total

1.  FEBio: finite elements for biomechanics.

Authors:  Steve A Maas; Benjamin J Ellis; Gerard A Ateshian; Jeffrey A Weiss
Journal:  J Biomech Eng       Date:  2012-01       Impact factor: 2.097

Review 2.  FEBio: History and Advances.

Authors:  Steve A Maas; Gerard A Ateshian; Jeffrey A Weiss
Journal:  Annu Rev Biomed Eng       Date:  2017-06-21       Impact factor: 9.590

Review 3.  Verification, validation and sensitivity studies in computational biomechanics.

Authors:  Andrew E Anderson; Benjamin J Ellis; Jeffrey A Weiss
Journal:  Comput Methods Biomech Biomed Engin       Date:  2007-06       Impact factor: 1.763

4.  Three-dimensional fibril-reinforced finite element model of articular cartilage.

Authors:  L P Li; J T M Cheung; W Herzog
Journal:  Med Biol Eng Comput       Date:  2009-03-06       Impact factor: 2.602

5.  Finite element algorithm for frictionless contact of porous permeable media under finite deformation and sliding.

Authors:  Gerard A Ateshian; Steve Maas; Jeffrey A Weiss
Journal:  J Biomech Eng       Date:  2010-06       Impact factor: 2.097

6.  New resource for the computation of cartilage biphasic material properties with the interpolant response surface method.

Authors:  Kathryn E Keenan; Lampros C Kourtis; Thor F Besier; Derek P Lindsey; Garry E Gold; Scott L Delp; Gary S Beaupre
Journal:  Comput Methods Biomech Biomed Engin       Date:  2009-08       Impact factor: 1.763

Review 7.  Nondestructive Techniques to Evaluate the Characteristics and Development of Engineered Cartilage.

Authors:  Joseph M Mansour; Zhenghong Lee; Jean F Welter
Journal:  Ann Biomed Eng       Date:  2016-01-27       Impact factor: 3.934

8.  Near-infrared (NIR) spectroscopy. A new method for arthroscopic evaluation of low grade degenerated cartilage lesions. Results of a pilot study.

Authors:  Gunter Spahn; Holger Plettenberg; Enrico Kahl; Hans M Klinger; Thomas Mückley; Gunther O Hofmann
Journal:  BMC Musculoskelet Disord       Date:  2007-05-29       Impact factor: 2.362

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.