A 16-year study of the microgap between 272 human titanium implants and their abutments.

Antonio Scarano1, Bartolomeo Assenza, Maurizio Piattelli, Giovanna Iezzi, Giulio C Leghissa, Alessandro Quaranta, Pietro Tortora, Adriano Piattelli.   

Abstract

A microgap has been described at the level of the implant-abutment connection. This microgap can be colonized by bacteria, and this fact could have relevance on the remodeling of the peri-implant crestal bone and on the long-term health of the peri-implant tissues. The authors report on 272 implants with screw- or cement-retained abutments retrieved from humans for different causes during a 16-year period. In the implants with screw-retained abutments, a 60-microm microgap was present at the level of implant-abutment connection. In some areas the titanium had sheared off from the surface and from the internal threads. The contact between the threads of the implant and those of the abutment was limited to a few areas. Bacteria were often present in the microgaps between implant and abutment and in the internal portion of the implants. In implants with cement-retained abutments, a 40-microm microgap was found at the level of the implant-abutment connection. No mechanical damage was observed at the level of the implant or of the abutment. All the internal voids were always completely filled by the cement. No bacteria were observed in the internal portion of the implants or at the level of the microgap. The differences in the size of the microgap between the two groups were statistically significant (P < .05). In conclusion, in screw-retained abutments the microgap can be a critical factor for colonization of bacteria, whereas in cement-retained abutments all the internal spaces were filled by cement. In these retrieved implants, the size of the microgap was markedly variable and much larger than that observed in vitro.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16447899     DOI: 10.1563/753.1

Source DB:  PubMed          Journal:  J Oral Implantol        ISSN: 0160-6972            Impact factor:   1.779


  17 in total

1.  Effects of multiple implantations of titanium healing abutments: Surface characteristics and microbial colonization.

Authors:  Sanjana S Jain; Sareda T J Schramm; Danyal A Siddiqui; Wenwen Huo; Kelli L Palmer; Thomas G Wilson; Danieli C Rodrigues
Journal:  Dent Mater       Date:  2020-06-24       Impact factor: 5.304

2.  Bacterial diversity of periodontal and implant-related sites detected by the DNA Checkerboard method.

Authors:  C do Nascimento; N Monesi; I Y Ito; J P M Issa; R F de Albuquerque Junior
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2011-04-30       Impact factor: 3.267

3.  Corrosion behavior of pure titanium in the presence of Actinomyces naeslundii.

Authors:  Song-Mei Zhang; Jing Qiu; Fei Tian; Xiao-Kui Guo; Fu-Qiang Zhang; Qing-Feng Huang
Journal:  J Mater Sci Mater Med       Date:  2013-02-21       Impact factor: 3.896

4.  Microgap and microleakage of a hybrid connection platform-switched implant system in the absence or presence of a silicone-based sealing agent.

Authors:  Pedro Ferrás Fernandes; Liliana Grenho; Maria Helena Fernandes; João Carlos Sampaio-Fernandes; Pedro Sousa Gomes
Journal:  Odontology       Date:  2021-08-27       Impact factor: 2.634

Review 5.  Is the internal connection more efficient than external connection in mechanical, biological, and esthetical point of views? A systematic review.

Authors:  Marcelo Coelho Goiato; Eduardo Piza Pellizzer; Emily Vivianne Freitas da Silva; Liliane da Rocha Bonatto; Daniela Micheline dos Santos
Journal:  Oral Maxillofac Surg       Date:  2015-04-25

6.  Factors determining the retentiveness of luting agents used with metal- and ceramic-based implant components.

Authors:  Cornelia Schiessl; Lina Schaefer; Christian Winter; Jan Fuerst; Martin Rosentritt; Florian Zeman; Michael Behr
Journal:  Clin Oral Investig       Date:  2012-07-31       Impact factor: 3.573

7.  Micro-leakage at the implant-abutment interface with different tightening torques in vitro.

Authors:  João Paulo da Silva-Neto; Marcel Santana Prudente; Thiago de Almeida Prado Naves Carneiro; Mauro Antônio de Arruda Nóbilo; Mario Paulo Amante Penatti; Flávio Domingues das Neves
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

8.  Influence of the implant-abutment connection design and diameter on the screw joint stability.

Authors:  Hyon-Mo Shin; Jung-Bo Huh; Mi-Jeong Yun; Young-Chan Jeon; Brian Myung Chang; Chang-Mo Jeong
Journal:  J Adv Prosthodont       Date:  2014-04-22       Impact factor: 1.904

9.  Physicochemical and microscopic characterization of implant-abutment joints.

Authors:  Patricia A Lopes; Adriana F P Carreiro; Rubens M Nascimento; Brendan R Vahey; Bruno Henriques; Júlio C M Souza
Journal:  Eur J Dent       Date:  2018 Jan-Mar

10.  Thermal conduction of titanium implants under CO2 laser irradiation in vitro.

Authors:  J Thomas Lambrecht; Tino Nyffeler; Manuela Linder
Journal:  Ann Maxillofac Surg       Date:  2012-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.