Intrapatient reproducibility of the BISxp monitor.

Dagmar J Niedhart1, Heiko A Kaiser, Eric Jacobsohn, Charles B Hantler, Alex S Evers, Michael S Avidan.   

Abstract

BACKGROUND: The Bispectral Index (BIS) reportedly reflects anesthetic depth. It is recommended that anesthetic agents should be titrated to maintain the BIS between 40 and 60 arbitrary BIS units during anesthesia. For anesthesia providers to follow this recommendation, the monitor should be predictably affected by different anesthetic agents and have good interpatient and intrapatient reproducibility. The authors hypothesized that when two BISxp devices (Aspect Medical Systems, Newton, MA) are placed concurrently on the same patient, their readings are concordant throughout the anesthetic period.
METHODS: Simultaneous BIS recordings from two BISxp monitors were obtained during anesthesia at 5-s intervals from 12 participants.
RESULTS: In total 22,860 concurrent paired BIS readings were obtained. For 10.7% of the time, there were sustained periods of 30 s or greater where the readings suggested a different depth of anesthesia. For 6% of the time, there were sustained periods of 30 s or greater where the readings differed by 10 or more arbitrary BIS units. The regression coefficient (R) for the two devices was 0.65 (range, 0.35-0.92). There was zero bias between the devices, and the 95% limits of agreement ranged between -18 and +17.
CONCLUSION: A conflicting anesthetic management was suggested by the simultaneous BIS readings 10.7% of the time. These results suggest that BISxp does not always provide a reproducible single number. Anesthesia providers should not rely exclusively on the BIS reading when assessing depth of anesthesia.

Entities:  

Mesh:

Year:  2006        PMID: 16436841     DOI: 10.1097/00000542-200602000-00007

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  9 in total

1.  NREM sleep staging using WAV(CNS) index.

Authors:  Gracee Agrawal; Mohammad Modarres; Tatjana Zikov; Stephane Bibian
Journal:  J Clin Monit Comput       Date:  2011-07-21       Impact factor: 2.502

Review 2.  Brain monitoring with electroencephalography and the electroencephalogram-derived bispectral index during cardiac surgery.

Authors:  Miklos D Kertai; Elizabeth L Whitlock; Michael S Avidan
Journal:  Anesth Analg       Date:  2012-01-17       Impact factor: 5.108

3.  Recent advance in patient monitoring.

Authors:  Tomoki Nishiyama
Journal:  Korean J Anesthesiol       Date:  2010-09-20

4.  Chaos analysis of EEG during isoflurane-induced loss of righting in rats.

Authors:  M B MacIver; Brian H Bland
Journal:  Front Syst Neurosci       Date:  2014-10-16

5.  Neurally adjusted ventilatory assist feasibility during anaesthesia: A randomised crossover study of two anaesthetics in a large animal model.

Authors:  Francesca Campoccia Jalde; Fredrik Jalde; Peter V Sackey; Peter J Radell; Staffan Eksborg; Mats K E B Wallin
Journal:  Eur J Anaesthesiol       Date:  2016-04       Impact factor: 4.330

6.  The bilateral bispectral and the composite variability indexes during anesthesia for unilateral surgical procedure.

Authors:  Pedro Lopes-Pimentel; Maylin Koo; Javier Bocos; Antoni Sabaté
Journal:  Saudi J Anaesth       Date:  2017 Jan-Mar

7.  Influence of electrocautery-induced electromagnetic interference on quantitative electroencephalographic monitoring of hypnosis during general anesthesia: comparison between the ADMS® and the BIS VISTATM.

Authors:  Dae-Hee Kim; Ji-Young Yoo; Jong-Yeop Kim; Soo-Hwan Ahn; Seongsu Kim; Sang-Kee Min
Journal:  Korean J Anesthesiol       Date:  2018-04-24

8.  Comparative evaluation of the bispectral index (BIS) and BISpro during propofol anaesthesia.

Authors:  Zhitao Li; Jiachun Cai; Jiexiong Li; Xianghui Xu; Limin Zheng
Journal:  J Int Med Res       Date:  2021-04       Impact factor: 1.671

9.  Variations in Values of State, Response Entropy and Haemodynamic Parameters Associated with Development of Different Epileptiform Patterns during Volatile Induction of General Anaesthesia with Two Different Anaesthetic Regimens Using Sevoflurane in Comparison with Intravenous Induct: A Comparative Study.

Authors:  Michał Stasiowski; Anna Duława; Izabela Szumera; Radosław Marciniak; Ewa Niewiadomska; Wojciech Kaspera; Lech Krawczyk; Piotr Ładziński; Beniamin Oskar Grabarek; Przemysław Jałowiecki
Journal:  Brain Sci       Date:  2020-06-12
  9 in total