Biomechanical evaluation of a dynamic pedicle screw fixation device.

Hua-Zi Xu1, Xiang-Yang Wang, Yong-Long Chi, Qing-An Zhu, Yan Lin, Qi-Shan Huang, Li-Yang Dai.   

Abstract

BACKGROUND: Recent innovations in dynamic devices have promised a reduction in stress shielding, protection of adjacent segment degeneration, and decreased implant failure. However, there have been few studies comparing the biomechanical properties of a rigid device in comparison to a dynamic posterior fixation device. The purpose of this study was to compare the immediate stability of a new dynamic pedicle screw fixation device with an equivalent rigid device.
METHODS: Six thoracolumbar cadaver spines (T10-L4) were fixed in a biomechanical testing frame. Pure moments of 10Nm were loaded in six directions: flexion, extension, right and left lateral bending, and right and left axial rotation. For each spine, four different stages were tested: intact, destabilization of the middle segment, fixation with the dynamic device, and fixation with the rigid device. Ranges of motion were measured using stereophotogrammetry. The specimens with each device were then subjected to flexion-compression loading for five cycles on a MTS 858 Universal Testing Machine. The average stiffness of the last three cycles was recorded.
FINDINGS: Both dynamic and rigid devices were found to provide stability for the injured segment in flexion-extension and lateral bending. In axial rotation, the devices could restore the stability to levels similar to those in an intact spine. Results also indicated a slight increase in range of motion in flexion-extension and significant reduction in stiffness of flexion-compression with the dynamic device (P < 0.01), in comparison to the rigid device.
INTERPRETATION: The dynamic device offers a system that may alter favorably the movement and load transmission of a spinal motion segment without sacrificing construct stability.

Entities:  

Mesh:

Year:  2006        PMID: 16434133     DOI: 10.1016/j.clinbiomech.2005.12.004

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  11 in total

Review 1.  Role of lumbar interspinous distraction on the neural elements.

Authors:  Alex Alfieri; Roberto Gazzeri; Julian Prell; Christian Scheller; Jens Rachinger; Christian Strauss; Andreas Schwarz
Journal:  Neurosurg Rev       Date:  2012-05-02       Impact factor: 3.042

2.  [Spinal column: implants and revisions].

Authors:  S M Krieg; H S Meyer; B Meyer
Journal:  Chirurg       Date:  2016-03       Impact factor: 0.955

3.  Characterization of the behavior of a novel low-stiffness posterior spinal implant under anterior shear loading on a degenerative spinal model.

Authors:  Angela D Melnyk; Jason D Chak; Vaneet Singh; Adrienne Kelly; Peter A Cripton; Charles G Fisher; Marcel F Dvorak; Thomas R Oxland
Journal:  Eur Spine J       Date:  2015-01-06       Impact factor: 3.134

4.  Effect of Device Rigidity and Physiological Loading on Spinal Kinematics after Dynamic Stabilization : An In-Vitro Biomechanical Study.

Authors:  Kwonsoo Chun; Inchul Yang; Namhoon Kim; Dosang Cho
Journal:  J Korean Neurosurg Soc       Date:  2015-11-30

5.  Stress Reduction in Adjacent Level Discs via Dynamic Instrumentation: A Finite Element Analysis.

Authors:  Antonio E Castellvi; Hao Huang; Tov Vestgaarden; Sunil Saigal; Deborah H Clabeaux; David Pienkowski
Journal:  SAS J       Date:  2007-05-01

6.  Biomechanical evaluation of pedicle screw-based dynamic stabilization devices for the lumbar spine: a systematic review.

Authors:  Cédric Y Barrey; Ravi K Ponnappan; Jason Song; Alexander R Vaccaro
Journal:  SAS J       Date:  2008-12-01

7.  Clinical outcomes of degenerative lumbar spinal stenosis treated with lumbar decompression and the Cosmic "semi-rigid" posterior system.

Authors:  Tuncay Kaner; Mehdi Sasani; Tunc Oktenoglu; Ahmet Levent Aydin; Ali Fahir Ozer
Journal:  SAS J       Date:  2010-12-01

8.  Dynamic stabilization for challenging lumbar degenerative diseases of the spine: a review of the literature.

Authors:  Tuncay Kaner; Ali Fahir Ozer
Journal:  Adv Orthop       Date:  2013-04-15

Review 9.  Controversies about interspinous process devices in the treatment of degenerative lumbar spine diseases: past, present, and future.

Authors:  Roberto Gazzeri; Marcelo Galarza; Alex Alfieri
Journal:  Biomed Res Int       Date:  2014-04-13       Impact factor: 3.411

10.  The Mechanical Effect of Rod Contouring on Rod-Screw System Strength in Spine Fixation.

Authors:  Nihat Acar; Ahmet Karakasli; Ahmet A Karaarslan; Mehmet Hilal Ozcanhan; Fatih Ertem; Mehmet Erduran
Journal:  J Korean Neurosurg Soc       Date:  2016-09-08