Masahito Tanaka1, Kazutoshi Yagi-Watanabe, Toru Yamada, Fusae Kaneko, Kazumichi Nakagawa.   

Abstract

We have developed an improved circular dichroism (CD) and linear dichroism (LD) simultaneous measurement system for the vacuum ultraviolet (VUV) region by polarization modulation techniques using a four-period Onuki-type crossed undulator as a polarized light source. The system has been constructed at the VUV beamline BL-5 in the electron storage ring TERAS, at AIST. Our improvements, in particular the adoption of an optical chopper as the detection method of incident light, have resulted in a flat baseline and a consequent simplification of the Mueller matrix calculation for our optical system. Based on the Mueller matrix calculation, we have successfully measured real VUV-CD and LD spectra of leucine films for wavelengths down to 160 nm with absolute optical constants. The obtained spectra show good consistency with spectra measured by conventional methods. Copyright 2006 Wiley-Liss, Inc.

Entities:  

Year:  2006        PMID: 16432920     DOI: 10.1002/chir.20234

Source DB:  PubMed          Journal:  Chirality        ISSN: 0899-0042            Impact factor:   2.437


  1 in total

1.  Chirality emergence in thin solid films of amino acids by polarized light from synchrotron radiation and free electron laser.

Authors:  Jun-Ichi Takahashi; Hiroyuki Shinojima; Michiko Seyama; Yuko Ueno; Takeo Kaneko; Kensei Kobayashi; Hajime Mita; Mashahiro Adachi; Masahito Hosaka; Masahiro Katoh
Journal:  Int J Mol Sci       Date:  2009-07-07       Impact factor: 6.208

  1 in total