Literature DB >> 16430798

Modified nasal cannula for simultaneous oxygen delivery and end-tidal CO2 monitoring during spontaneous breathing.

F Yanagidate1, S Dohi.   

Abstract

BACKGROUND AND OBJECTIVES: Supplemental oxygen is commonly given via nasal cannulae in spontaneously breathing patients. Our modified nasal cannula with a clamp between the nasal prongs can provide O2 via one nostril and CO2 can be sampled through the other one. We have studied whether this cannula can provide oxygenation similar to a standard cannula without affecting end-tidal CO2 monitoring.
METHODS: Eighty-six patients were studied during spinal anaesthesia and sedation. In 15 patients, arterial blood was sampled while O2 was delivered at flow rates of 0, 2 and 4 L min(-1), with or without clamping between the prongs of our modified nasal cannula. In the remaining 71 patients, arterial O2 was measured while using our modified nasal cannula with the clamp applied. End-tidal CO2 was recorded on a capnograph and the correlation between end-tidal and arterial values with our modified nasal cannula was investigated.
RESULTS: No end-tidal CO2 waveforms were found with oxygen flow greater than 2L min(-1) without clamping between the prongs. With clamping there was a significant correlation (r = 0.83) between arterial and end-tidal CO2. A Bland-Altman analysis revealed a bias of 0.49 kPa with precision of +/-0.76 kPa. Arterial oxygenation was not affected by our modified nasal prongs with clamp as compared to the standard cannula.
CONCLUSION: Our modified nasal cannula can provide continuous monitoring of end-tidal CO2 without affecting oxygen delivery in sedated, spontaneously breathing patients.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16430798     DOI: 10.1017/S0265021505002279

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  6 in total

1.  Oral capnography is more effective than nasal capnography during sedative upper gastrointestinal endoscopy.

Authors:  Wei-Nung Teng; Chien-Kun Ting; Yu-Tzu Wang; Ming-Chih Hou; Mei-Yung Tsou; Huihua Chiang; Chun-Li Lin
Journal:  J Clin Monit Comput       Date:  2017-05-24       Impact factor: 2.502

2.  Evaluation and application of a method for estimating nasal end-tidal O2 fraction while administering supplemental O2.

Authors:  Kyle M Burk; Kai Kuck; Joseph A Orr
Journal:  J Clin Monit Comput       Date:  2019-02-06       Impact factor: 2.502

3.  A Novel Real-time Carbon Dioxide Analyzer for Health and Environmental Applications.

Authors:  Di Zhao; Dylan Miller; Xiaojun Xian; Francis Tsow; Erica S Forzani
Journal:  Sens Actuators B Chem       Date:  2014-05-01       Impact factor: 7.460

4.  Is it safe to use propofol in the emergency department? A randomized controlled trial to compare propofol and midazolam.

Authors:  Nik Hisamuddin Nik Ab Rahman; Ahmad Hashim
Journal:  Int J Emerg Med       Date:  2010-03-25

5.  Comparison of the effects of ketamine and fentanyl-midazolam-medetomidine for sedation of rhesus macaques (Macaca mulatta).

Authors:  Henri G M J Bertrand; Yvette C Ellen; Stevie O'Keefe; Paul A Flecknell
Journal:  BMC Vet Res       Date:  2016-06-08       Impact factor: 2.741

6.  Evaluation of the Integrated Pulmonary Index® during non-anesthesiologist sedation for percutaneous endoscopic gastrostomy.

Authors:  Florian Alexander Michael; Jan Peveling-Oberhag; Eva Herrmann; Stefan Zeuzem; Jörg Bojunga; Mireen Friedrich-Rust
Journal:  J Clin Monit Comput       Date:  2020-07-30       Impact factor: 2.502

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.