Design and validation of an analyser to measure sulphur hexafluoride gas during respiration.

J K K Kanhai1, F Eijskoot, E G J Reinders, H A Bruining, G J Puppels.   

Abstract

The study presents the results of the development of an analyser to measure sulphur hexafluoride (SF6) gas in breathing circuits, for application is studies of lung function. The analyser consists of an in-line breathing circuit measurement transducer and a compact unit for signal treatment. The detector unit of the analyser consists of a near-infrared light source, a bandpass filter and a pyro-electrical detector. When incremental steps of SF6 gas between 0 and 2% were presented to the analyser, the maximum deviation from the theoretical calibration curve was calculated to be 0.01% SF6. The step response of the analyser (10-90%) was 250 ms. The sensitivity of the analyser to ambient temperature was 0.01% SF6 degrees C(-1) in the range between 0 and 2% SF6. It is concluded that the analyser presented is accurate, and has a sufficient response speed to be used in clinical measurement settings. Furthermore, the analyser is resistant to changes in temperature, gas flow, orientation and movement, which are likely to occur in clinical measurement settings.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16411643     DOI: 10.1007/BF02351044

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   3.079


  22 in total

Review 1.  [FRC measurement in intensive care patients. A definition of standards].

Authors:  H J Wauer; B A Lorenz; W J Kox
Journal:  Anaesthesist       Date:  1998-10       Impact factor: 1.041

2.  Assessment of the Hewlett-Packard HP47210A capnometer.

Authors:  S M Kinsella
Journal:  Br J Anaesth       Date:  1985-09       Impact factor: 9.166

3.  Automated measurement of functional residual capacity by sulfur hexafluoride washout.

Authors:  T D East; K P Andriano; N L Pace
Journal:  J Clin Monit       Date:  1987-01

4.  Changes in functional residual capacity of the lung after operation.

Authors:  J R Meyers; L Lembeck; H O'Kane; A E Baue
Journal:  Arch Surg       Date:  1975-05

5.  An analyzer for in-line measurement of expiratory sulfur hexafluoride concentration.

Authors:  C Jonmarker; R Castor; B Drefeldt; O Werner
Journal:  Anesthesiology       Date:  1985-07       Impact factor: 7.892

6.  Computer-controlled positive end-expiratory pressure titration for effective oxygenation without frequent blood gases.

Authors:  T D East; J C in't Veen; T A Jonker; N L Pace; S McJames
Journal:  Crit Care Med       Date:  1988-03       Impact factor: 7.598

7.  Measurement of functional residual capacity by sulfur hexafluoride in small-volume lungs during spontaneous breathing and mechanical ventilation.

Authors:  A Schulze; P Schaller; A Töpfer; H Kirpalani
Journal:  Pediatr Res       Date:  1994-04       Impact factor: 3.756

8.  Measurement of lung volume by sulfur hexafluoride washout during spontaneous and controlled ventilation: further development of a method.

Authors:  A Larsson; D Linnarsson; C Jonmarker; B Jonson; H Larsson; O Werner
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

9.  Gas mixing in the lungs of dogs and pigs.

Authors:  S Kelly; C Cohen; E Powell; M Paiva; L A Engel
Journal:  Respir Physiol       Date:  1982-03

10.  A fast, digitally controlled flow proportional gas injection system for studies in lung function.

Authors:  J K K Kanhai; P J Caspers; E G J Reinders; J C Pompe; H A Bruining; G J Puppels
Journal:  IEEE Trans Biomed Eng       Date:  2003-11       Impact factor: 4.538

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.