Optimizing T2-weighted magnetic resonance sequences for surface coil microimaging of the eye with regard to lid, eyeball and head moving artifacts.

Takayuki Obata1, Koji Uemura, Hiroi Nonaka, Mitsuru Tamura, Shuji Tanada, Hiroo Ikehira.   

Abstract

PURPOSE: To acquire high-resolution magnetic resonance (MR) images, we developed a new blinking artifact reduced pulse (BARP) sequence with a surface coil specialized for microscopic imaging (47 mm in diameter).
MATERIALS AND METHODS: To reduce eye movement, we ascertained that the subjects' eyes were kept open and fixated to the target in the 1.5-T MR gantry. To reduce motion artifacts from blinking, we inserted rest periods for blinking (1.5 s within every 5 s) during MR scanning (T2-weighted fast spin echo; repetition time, 5 s; echo time, 100 ms; echo train, 11; matrix, 256 x 128; field of view, 5 cm; 1-mm thickness x 30 slices). Three scans (100 s x 3) were performed for each normal subject, and they were added together after automatic adjustment for location to reduce quality loss caused by head motion.
RESULTS: T2-weighted MR images were acquired with a high resolution and a high signal-to-noise ratio. Motion artifacts were reduced with BARP, as compared with those with random blinking. Intraocular structures such as the iris and ciliary muscles were clearly visualized. Because the whole eye can be covered with a 1-mm thickness by this method, three-dimensional maps can easily be generated from the obtained images.
CONCLUSION: The application of BARP with a surface coil of the human eye might become a useful and widely adopted procedure for MR microimaging.

Entities:  

Mesh:

Year:  2005        PMID: 16410184     DOI: 10.1016/j.mri.2005.10.011

Source DB:  PubMed          Journal:  Magn Reson Imaging        ISSN: 0730-725X            Impact factor:   2.546


  3 in total

1.  Clinical applications of high-resolution ocular magnetic resonance imaging.

Authors:  Keizo Tanitame; Takashi Sone; Yoshiaki Kiuchi; Kazuo Awai
Journal:  Jpn J Radiol       Date:  2012-08-28       Impact factor: 2.374

2.  Uveal melanoma: evaluation of extrascleral extension using thin-section MR of the eye with surface coils.

Authors:  Tommaso Tartaglione; Monica Maria Pagliara; Mariacarmela Sciandra; Carmela Grazia Caputo; Rosalinda Calandrelli; Gina Fabrizi; Simona Gaudino; Maria Antonietta Blasi; Cesare Colosimo
Journal:  Radiol Med       Date:  2014-01-28       Impact factor: 3.469

3.  Ocular volumetry using fast high-resolution MRI during visual fixation.

Authors:  K Tanitame; T Sone; T Miyoshi; N Tanitame; K Otani; Y Akiyama; M Takasu; S Date; Y Kiuchi; K Awai
Journal:  AJNR Am J Neuroradiol       Date:  2012-10-04       Impact factor: 3.825

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.