Literature DB >> 16409572

On-line adaptive algorithm with glucose prediction capacity for subcutaneous closed loop control of glucose: evaluation under fasting conditions in patients with Type 1 diabetes.

H C Schaller1, L Schaupp, M Bodenlenz, M E Wilinska, L J Chassin, P Wach, T Vering, R Hovorka, T R Pieber.   

Abstract

AIMS: To evaluate an algorithm with glucose prediction capacity and continuous adaptation of patient parameters-a model predictive control (MPC) algorithm-to control blood glucose concentration during fasting conditions in patients with Type 1 diabetes. In the subcutaneous (sc) route within a closed loop system.
METHODS: Paired experiments were performed in six patients. Over 8 h the MPC algorithm was used to control glucose with s.c. insulin administration and two different glucose monitoring protocols: first, the algorithm was provided with intravenous (i.v.) glucose values for insulin dosage calculation directly (i.v.-s.c. route). Then, in the second experiment, i.v. glucose values were fed to the MPC with a delay of 30 min to simulate s.c. glucose measurements ('s.c.'-s.c. route). In both experiments plasma glucose, insulin dosage, and serum insulin levels were analysed.
RESULTS: Glucose concentration was brought from hyper- to normoglycaemia and kept in the physiological range (6-7 mmol/l) with both routes in all subjects. Mean glucose concentration reached the threshold of 7 mmol/l approximately 2 (i.v.-s.c. route) and 3 ('s.c.'-s.c. route) hours after the start of glucose control with the MPC. During the last 2 h of automated glucose control, mean glucose concentration was 6.3 +/- 0.2 mmol/l and 6.6 +/- 0.3 mmol/l for i.v.-s.c. and 's.c.'-s.c. route, respectively. Glucose concentration, insulin doses, and serum insulin levels did not differ significantly between routes (P > 0.05).
CONCLUSIONS: The MPC algorithm is suitable for glucose control during fasting within an extracorporeal artificial beta-cell in the subcutaneous route Type 1 diabetic patients.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16409572     DOI: 10.1111/j.1464-5491.2006.01695.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  19 in total

1.  AP@home: a novel European approach to bring the artificial pancreas home.

Authors:  Lutz Heinemann; Carsten Benesch; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2011-11-01

2.  Postprandial glycemic excursions with the use of a closed-loop platform in subjects with type 1 diabetes: a pilot study.

Authors:  Arianne C van Bon; Jeroen Hermanides; Robin Koops; Joost B L Hoekstra; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2010-07-01

Review 3.  Toward closing the loop: an update on insulin pumps and continuous glucose monitoring systems.

Authors:  Tandy Aye; Jen Block; Bruce Buckingham
Journal:  Endocrinol Metab Clin North Am       Date:  2010-09       Impact factor: 4.741

4.  A benchtop closed-loop system controlled by a bio-inspired silicon implementation of the pancreatic beta cell.

Authors:  Nick Oliver; Pantelis Georgiou; Desmond Johnston; Christofer Toumazou
Journal:  J Diabetes Sci Technol       Date:  2009-11-01

Review 5.  A Review of Emerging Technologies for the Management of Diabetes Mellitus.

Authors:  Konstantia Zarkogianni; Eleni Litsa; Konstantinos Mitsis; Po-Yen Wu; Chanchala D Kaddi; Chih-Wen Cheng; May D Wang; Konstantina S Nikita
Journal:  IEEE Trans Biomed Eng       Date:  2015-08-19       Impact factor: 4.538

6.  Automatic data processing to achieve a safe telemedical artificial pancreas.

Authors:  M Elena Hernando; Gema García-Sáez; Iñaki Martínez-Sarriegui; Agustín Rodríguez-Herrero; Carmen Pérez-Gandía; Mercedes Rigla; Alberto de Leiva; Ismael Capel; Belén Pons; Enrique J Gómez
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

Review 7.  Wearable and implantable pancreas substitutes.

Authors:  Leonardo Ricotti; Tareq Assaf; Paolo Dario; Arianna Menciassi
Journal:  J Artif Organs       Date:  2012-09-20       Impact factor: 1.731

8.  The artificial pancreas: how sweet engineering will solve bitter problems.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2007-01

9.  A Run-to-Run Control Strategy to Adjust Basal Insulin Infusion Rates in Type 1 Diabetes.

Authors:  Cesar C Palerm; Howard Zisser; Lois Jovanovič; Francis J Doyle
Journal:  J Process Control       Date:  2008       Impact factor: 3.666

10.  An integrated multivariable artificial pancreas control system.

Authors:  Kamuran Turksoy; Lauretta T Quinn; Elizabeth Littlejohn; Ali Cinar
Journal:  J Diabetes Sci Technol       Date:  2014-04-07