Infection risk associated with a closed luer access device.

D Adams1, T Karpanen, T Worthington, P Lambert, T S J Elliott.   

Abstract

The potential for microbial contamination associated with a recently developed needleless closed luer access device (CLAD) (Q-Syte; Becton Dickinson, Sandy, UT, USA) was evaluated in vitro. Compression seals of 50 multiply activated Q-Syte devices were inoculated with Staphylococcus epidermidis NCTC 9865 in 25% (v/v) human blood and then disinfected with 70% (v/v) isopropyl alcohol followed by flushing with 0.9% (w/v) sterile saline. Forty-eight of 50 (96%) saline flushes passed through devices that had been activated up to a maximum of 70 times remained sterile. A further 25 Q-Syte CLADs that had undergone multiple activations were challenged with prefilled 0.9% (w/v) sterile saline syringes, the external luer tips of which had been inoculated with S. epidermidis NCTC 9865 prior to accessing the devices. None of the devices that had been accessed up to 70 times allowed passage of micro-organisms, despite challenge micro-organisms being detected on both the syringe tip after activation and the compression seals before decontamination. These findings suggest that the Q-Syte CLAD may be activated up to 70 times with no increased risk of microbial contamination within the fluid pathway. The device may also offer protection from the external surface of syringe tips contaminated with micro-organisms.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16406139     DOI: 10.1016/j.jhin.2005.09.016

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  3 in total

1.  In vitro study on the disinfectability of two split-septum needle-free connection devices using different disinfection procedures.

Authors:  Steffen Engelhart; Martin Exner; Arne Simon
Journal:  GMS Hyg Infect Control       Date:  2015-12-09

2.  How much is red blood cell fragmentation increased by the use of closed luer lock access devices on catheter hubs?

Authors:  Sunny Eloot; Jean-Yves De Vos; Remi Hombrouckx
Journal:  NDT Plus       Date:  2008-09-01

3.  Impact of a modified Broviac maintenance care bundle on bloodstream infections in paediatric cancer patients.

Authors:  Rhoikos Furtwängler; Carolin Laux; Norbert Graf; Arne Simon
Journal:  GMS Hyg Infect Control       Date:  2015-11-16
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.