Experimental evaluation of errors in the measurement of respiratory parameters of the newborn performed by a continuous flow neonatal ventilator.

P Cappa1, S A Sciuto, S Silvestri.   

Abstract

Pulmonary ventilators for intensive care provide information on, among many other patient respiratory parameters, patient resistance, compliance and 'work of breathing' values calculated from pressure and flow data patterns according to a widely utilized algorithm. The effects induced by the breathing circuit and analogue filtering of the ventilator measuring system are experimentally investigated during controlled ventilation. Three main phenomena are observed: (a) errors in calculation of resistance and compliance due to filtering of pressure and flow waveforms; (b) the presence of pressure oscillations at the beginning of inspiration and expiration phases; and (c) the phase shift between pressure and flow waveforms. The experimental evaluation of the measuring system of a neonatal ventilator is then conducted and: (a) a delay in pressure and flow measurement synchronization equal to 22 +/- 2 ms is evaluated; moreover, (b) a difference between the values provided by the ventilator and those measured by the reference experimental setup on respiratory parameters such as the compliance, resistance and work of breathing that lies in the range of 7-16% of reading is observed.

Entities:  

Mesh:

Year:  2006        PMID: 16393851     DOI: 10.1080/03091900500040295

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  4 in total

1.  Precision of continuous neonatal ventilator respiratory mechanics is improved with selected optimal respiratory cycles.

Authors:  Vincent Rigo; Estelle Graas; Jacques Rigo
Journal:  Eur J Pediatr       Date:  2011-12-06       Impact factor: 3.183

2.  Technical problems with dynamic compliance evaluation in neonates and infants.

Authors:  Daniele De Luca; Marco Piastra; Giorgio Conti
Journal:  Intensive Care Med       Date:  2012-03-17       Impact factor: 17.440

3.  Secretory phospholipase A2 and neonatal respiratory distress: pilot study on broncho-alveolar lavage.

Authors:  Daniele De Luca; Silvia Baroni; Giovanni Vento; Marco Piastra; Domenico Pietrini; Federica Romitelli; Ettore Capoluongo; Costantino Romagnoli; Giorgio Conti; Enrico Zecca
Journal:  Intensive Care Med       Date:  2008-07-22       Impact factor: 17.440

4.  In Memoriam: Paolo Cappa.

Authors:  Eduardo Palermo; Stefano Rossi; Fabrizio Patanè; Jeffrey Laut; Maurizio Porfiri
Journal:  Sensors (Basel)       Date:  2017-11-18       Impact factor: 3.576

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.