Literature DB >> 16390564

Optimal depth of insertion of left-sided double-lumen endobronchial tubes cannot be predicted from body height in below average-sized adult patients.

M Yasumoto1, K Higa, K Nitahara, S Shono, T Hamada.   

Abstract

BACKGROUND AND
OBJECTIVE: The optimal depth of insertion of left-sided double-lumen endobronchial tubes is strongly correlated with body height in average-sized adults. However, this relationship has not been studied in below average-sized adult patients. We investigated whether or not there is a clinically useful relationship in below average-sized adult patients.
METHODS: One hundred and ninety six consecutive adult patients undergoing thoracic surgery under one-lung anaesthesia (body height < or = 155 cm) were included in this study. Left-sided double-lumen tubes were inserted under the guidance of a fibre-optic bronchoscope. Optimal depth was defined as the proximal surface of the bronchial cuff positioned just below the carina.
RESULTS: There was a statistically significant positive correlation between body height and the optimal depth of insertion (r = 0.61, P < 0.0001); however, the correlation coefficient was low. The actual optimal depth of insertion of one patient was even 4.5 cm shorter than that obtained from the equation.
CONCLUSION: Although there was a statistically significant correlation between body height and the optimal depth of insertion of left sided double lumen tubes in adult patients of short stature (< or = 155 cm), clinical application of the equation is not warranted and these tubes should be inserted under direct vision with a fibre-optic bronchoscope.

Entities:  

Mesh:

Year:  2006        PMID: 16390564     DOI: 10.1017/S0265021505001742

Source DB:  PubMed          Journal:  Eur J Anaesthesiol        ISSN: 0265-0215            Impact factor:   4.330


  3 in total

Review 1.  A practical approach to adult one-lung ventilation.

Authors:  V Ashok; J Francis
Journal:  BJA Educ       Date:  2017-12-06

2.  Efficacy of a New Blind Insertion Technique of Arndt Endobronchial Blocker for Lung Isolation: Comparison With Conventional Bronchoscope-Guided Insertion Technique-A Pilot Study.

Authors:  Peng Liang; Juan Ni; Cheng Zhou; Hai Yu; Bin Liu
Journal:  Medicine (Baltimore)       Date:  2016-05       Impact factor: 1.889

3.  Insertion depth of left-sided double-lumen endobroncheal tube: A new predictive formula.

Authors:  Abdelazeem Eldawlatly; Ahmed Alqatari; Naveed Kanchi; Amir Marzouk
Journal:  Saudi J Anaesth       Date:  2019 Jul-Sep
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.