Derived 12-lead electrocardiogram in the assessment of ST-segment deviation and cardiac rhythm.

Darawan Chantad1, Rungroj Krittayaphong, Chulalak Komoltri.   

Abstract

BACKGROUND: There are little data on the validation of 12-lead electrocardiogram (ECG) derived by the EASI lead system used for continuous monitoring in critical care settings.
OBJECTIVE: The objectives of this study were to determine the accuracy of 12-lead ECG derived by the EASI lead system in the detection of ST-segment deviation and cardiac rhythm compared with the standard 12-lead ECG.
METHODS: All patients admitted to the coronary care unit were studied. Kappa statistics was used to calculate the agreement between both ECG systems in the determination of cardiac rhythm and premature ventricular complex morphology. ST-segment analysis was performed in patients with acute coronary syndromes. Pearson correlation was used to correlate the ST-segment deviation between both techniques. The sensitivity and specificity of the determination of significant ST-segment deviation by the EASI lead system were calculated.
RESULTS: There were a total of 282 patients enrolled in this study. There was a complete agreement in the interpretation of cardiac rhythm between the 2 methods (kappa = 1). Analysis of ST-segment deviation of 12-lead ECG also showed a significant correlation (correlation coefficient varied from 0.62 in lead I to 0.823 in lead aVF with a P value of <.001 in all leads) between the 2 methods with very high sensitivity and specificity in the detection of significant ST-segment elevation and depression.
CONCLUSION: The 12-lead ECG derived by the EASI lead system is an accurate and reliable information for the assessment of ST-segment deviation and cardiac rhythm in critically ill patients.

Entities:  

Mesh:

Year:  2005        PMID: 16387043     DOI: 10.1016/j.jelectrocard.2005.01.011

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  2 in total

Review 1.  Technical mistakes during the acquisition of the electrocardiogram.

Authors:  Javier García-Niebla; Pablo Llontop-García; Juan Ignacio Valle-Racero; Guillem Serra-Autonell; Velislav N Batchvarov; Antonio Bayés de Luna
Journal:  Ann Noninvasive Electrocardiol       Date:  2009-10       Impact factor: 1.468

2.  Electrode positions, transformation coordinates for ECG reconstruction from S-ICD vectors.

Authors:  David G Wilson; Peter L Cronbach; D Panfilo; Saul E Greenhut; Berthold P Stegemann; John M Morgan
Journal:  Data Brief       Date:  2017-02-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.