Glycated hemoglobin assessment in clinical practice: comparison of the A1cNow point-of-care device with central laboratory testing (GOAL A1C Study).

Laurence Kennedy1, William H Herman.   

Abstract

BACKGROUND: The Glycemic Optimization with Algorithms and Labs At Po1nt of Care (GOAL A1C) Study assessed the effect of titration monitoring strategies and methods of A1C testing on glycemic control in patients with type 2 diabetes failing oral therapy and beginning basal insulin glargine. The availability of both point-of-care (POC) and central laboratory A1C values provided an opportunity to evaluate correlation and statistical agreement between these methods of testing. This analysis forms the basis of the current report.
METHODS: This is a 24-week, randomized, four-arm, open-label study conducted in 7,758 subjects enrolled at 2,130 sites. At baseline, patients had A1C measurements both by POC testing using the A1cNow device (Metrika, Inc., Sunnyvale, CA), which applies an immunoassay method, and by central laboratory analysis using ion exchange high-performance liquid chromatography. These measures were compared statistically.
RESULTS: An r value of 0.72 was calculated for POC and laboratory A1C assessments. Although the mean POC A1C values were in agreement with the central laboratory values, there was a large range in individual POC A1C values.
CONCLUSIONS: POC testing of A1C in predominantly primary care settings using the A1cNow device was correlated with central laboratory results. The correlation was less than expected based on each method's reproducibility data. Although there was agreement between the average POC A1C values and the corresponding central laboratory values, the dispersion of individual POC A1C values was large. Thus, we conclude that these two methods of A1C testing should not be used interchangeably.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16386096     DOI: 10.1089/dia.2005.7.907

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  22 in total

1.  An information-motivation-behavioral skills analysis of diet and exercise behavior in Puerto Ricans with diabetes.

Authors:  Chandra Y Osborn; K Rivet Amico; William A Fisher; Leonard E Egede; Jeffrey D Fisher
Journal:  J Health Psychol       Date:  2010-05-07

2.  A brief culturally tailored intervention for Puerto Ricans with type 2 diabetes.

Authors:  Chandra Y Osborn; K R Amico; Noemi Cruz; Ann A O'Connell; Rafael Perez-Escamilla; Seth C Kalichman; Scott A Wolf; Jeffrey D Fisher
Journal:  Health Educ Behav       Date:  2010-11-12

3.  Relative inaccuracy of the A1cNow in children with type 1 diabetes.

Authors:  Larry Fox; Mariya Dontchev; Katrina Ruedy; Roy Beck; Craig Kollman; Laurel Messer; Julie Coffey; Darrell Wilson; Elizabeth Doyle; William Tamborlane; Michael Steffes
Journal:  Diabetes Care       Date:  2007-01       Impact factor: 19.112

4.  The ARMS-D out performs the SDSCA, but both are reliable, valid, and predict glycemic control.

Authors:  Lindsay S Mayberry; Jeffrey S Gonzalez; Kenneth A Wallston; Sunil Kripalani; Chandra Y Osborn
Journal:  Diabetes Res Clin Pract       Date:  2013-09-26       Impact factor: 5.602

5.  Improving outcomes with POCT for HbA1c and blood ketone testing.

Authors:  Lori Laffel
Journal:  J Diabetes Sci Technol       Date:  2007-01

6.  Assessment of the performance of A1CNow(+) and development of an error grid analysis graph for comparative hemoglobin A1c measurements.

Authors:  Fusong Jiang; Xuhong Hou; Jun Lu; Jian Zhou; Fengdi Lu; Kai Kan; Junling Tang; Yuqian Bao; Weiping Jia
Journal:  Diabetes Technol Ther       Date:  2014-04-25       Impact factor: 6.118

7.  Stressors may compromise medication adherence among adults with diabetes and low socioeconomic status.

Authors:  Chandra Y Osborn; Lindsay Satterwhite Mayberry; Julie A Wagner; Garry W Welch
Journal:  West J Nurs Res       Date:  2014-02-25       Impact factor: 1.967

8.  Family members' obstructive behaviors appear to be more harmful among adults with type 2 diabetes and limited health literacy.

Authors:  Lindsay S Mayberry; Russell L Rothman; Chandra Y Osborn
Journal:  J Health Commun       Date:  2014

9.  Assessment of glycated hemoglobin using A1CNow+ point-of-care device as compared to central laboratory testing.

Authors:  Justin R Arrendale; Sonia E Cherian; Issam Zineh; Mark J Chirico; James R Taylor
Journal:  J Diabetes Sci Technol       Date:  2008-09

10.  Liraglutide vs insulin glargine and placebo in combination with metformin and sulfonylurea therapy in type 2 diabetes mellitus (LEAD-5 met+SU): a randomised controlled trial.

Authors:  D Russell-Jones; A Vaag; O Schmitz; B K Sethi; N Lalic; S Antic; M Zdravkovic; G M Ravn; R Simó
Journal:  Diabetologia       Date:  2009-08-14       Impact factor: 10.122

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.