Comprehensive analysis of the behavior of an amperometric biosensor incorporated in a micro flow channel was conducted by changing the structural and conditional parameters. The device used in the characterization consisted of a thin-film three-electrode system and a silicone rubber flow channel. An enzyme, glucose oxidase, was immobilized either at the bottom of the silicone rubber flow channel or on the electrode substrate. The flow rate, concentration, position of the immobilized enzyme, and channel height were changed, and the changes in the output current and the conversion efficiency were examined. When the flow rate and/or the channel height decreased, the output current and the conversion efficiency significantly increased. The conversion efficiency also increased by decreasing the concentration. The tendency of the flow dependence was reversed when the position of the immobilized enzyme was changed from the silicone rubber side to the electrode substrate. In addition, the influence of l-ascorbic acid was reduced by placing additional working electrodes in the upper stream. l-Ascorbic acid was eliminated more effectively as the flow rate decreased and the area of the working electrode for elimination increased.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16384695     DOI: 10.1016/j.bios.2005.11.009

Source DB:  PubMed          Journal:  Biosens Bioelectron        ISSN: 0956-5663            Impact factor:   10.618


  2 in total

1.  Flow biosensing and sampling in indirect electrochemical detection.

Authors:  Francesco Lamberti; Camilla Luni; Alessandro Zambon; Pier Andrea Serra; Monica Giomo; Nicola Elvassore
Journal:  Biomicrofluidics       Date:  2012-04-20       Impact factor: 2.800

2.  TheClinical Research Tool: a high-performance microdialysis-based system for reliably measuring interstitial fluid glucose concentration.

Authors:  Gregor Ocvirk; Martin Hajnsek; Ralph Gillen; Arnfried Guenther; Gernot Hochmuth; Ulrike Kamecke; Karl-Heinz Koelker; Peter Kraemer; Karin Obermaier; Cornelia Reinheimer; Nina Jendrike; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2009-05-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.