Effective dose for patients undergoing coronary and femoral intravascular radiotherapy involving an HDR 192Ir source.

X George Xu1.   

Abstract

Effective dose equivalent (EDE), and effective dose (ED) for coronary and femoral Intravascular brachytherapy (IVBT) procedures involving a 370 GBq (10 Ci) HDR 192Ir gamma source are tabulated. MIRD stylised models and the MCNP Monte Carlo code were used for the calculations. For coronary irradiation, the normalised EDE is 0.18 mSv (GBq min)(-1) and the ED is 0.056 mSv (GBq min)(-1). For femoral IVBT, the normalised EDE is 0.01629 mSv (GBq min)(-1) and the ED is 0.01195 mSv (GBq min)(-1). Although the medical benefits to a patient undergoing IVBT are often significant and justified, patient doses are high compared with dose limits for radiation protection purposes. As IVBT is becoming a routine procedure, data in this paper could be useful to manage the procedures efficiently.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16381731     DOI: 10.1093/rpd/nci201

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  2 in total

1.  Comparison of organ doses for patients undergoing balloon brachytherapy of the breast with HDR 192Ir or electronic sources using monte carlo simulations in a heterogeneous human phantom.

Authors:  Matthew M Mille; X George Xu; Mark J Rivard
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

2.  Computed organ doses to an Indian reference adult during brachytherapy treatment of esophagus, breast, and neck cancers.

Authors:  Biju Keshavkumar
Journal:  J Med Phys       Date:  2012-07